5.4.3.3. Auto Range Mode (AUTO)

In AUTO range mode, the analyzer automatically switches the reporting range between
two user-defined ranges (low and high). The unit will switch from low range to high
range when either the SO

2

concentration exceeds 98% of the low range span. The unit

will return from high range back to low range once both the SO

2

concentration falls

below 75% of the low range span.
In AUTO Range mode the instrument reports the same data in the same range on both
the A1 and A2 outputs and automatically switches both outputs between ranges as
described above. Also, the RANGE test function displayed on the front panel will be
replaced by two separate functions:

RANGE1: The LOW range setting for all analog outputs.
RANGE2: The HIGH range setting for all analog outputs.

The high/low range status is also reported through the external, digital status bits (refer
to Section 8.1.1).
To set individual ranges press the following control button sequence.



SAMPLE RANGE = 500.000 PPB SO2 =XXX.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK MORE

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

SETUP X.X

RANGE MODE:

SNGL

SNGL IND

AUTO

ENTR EXIT

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

EXIT

x 2 returns

to the main

SAMPLE

display

SETUP X.X

RANGE MODE:

AUTO

SNGL

IND AUTO

ENTR

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET

UNIT

EXIT

SETUP X.X

HIGH RANGE:

500.0 Conc

0 0 5 0 0 .0

ENTR

EXIT

SETUP X.X

LOW RANGE:

500.0 Conc

0 0 5 0 0 .0

ENTR

EXIT

Toggle the numeral

buttons to set the

LOW

and HIGH

range value.

ENTR accepts the

new setting, EXIT

ignores the new

setting.

Figure 5-6: SETUP RNGE – Auto Range Mode

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

In AUTO range mode, the LOW and HIGH ranges have separate slopes
and offsets for computing SO

2

concentration. The two ranges must be

independently calibrated.

5.4.4. Range Units

The 6200T can display concentrations in parts per billion (10

9

mols per mol, PPB), parts

per million (10

6

mols per mol, PPM), micrograms per cubic meter (µg/m

3

, UGM) or