Range units, 6200t total sulfur analyzer setup menu, Auto range mode (auto) – Teledyne 6200T - Sulfides Analyzer User Manual
Page 103: Mols per mol, ppb), parts per million (10, Ugm) or, Teledyne analytical instruments 103, Figure 5-6: setup rnge – auto range mode
6200T Total Sulfur Analyzer
Setup Menu
Teledyne Analytical Instruments
103
5.4.3.3. Auto Range Mode (AUTO)
In AUTO range mode, the analyzer automatically switches the reporting range between
two user-defined ranges (low and high). The unit will switch from low range to high
range when either the SO
2
concentration exceeds 98% of the low range span. The unit
will return from high range back to low range once both the SO
2
concentration falls
below 75% of the low range span.
In AUTO Range mode the instrument reports the same data in the same range on both
the A1 and A2 outputs and automatically switches both outputs between ranges as
described above. Also, the RANGE test function displayed on the front panel will be
replaced by two separate functions:
RANGE1: The LOW range setting for all analog outputs.
RANGE2: The HIGH range setting for all analog outputs.
The high/low range status is also reported through the external, digital status bits (refer
to Section 8.1.1).
To set individual ranges press the following control button sequence.
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP X.X
RANGE MODE:
SNGL
SNGL IND
AUTO
ENTR EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
EXIT
x 2 returns
to the main
SAMPLE
display
SETUP X.X
RANGE MODE:
AUTO
SNGL
IND AUTO
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
HIGH RANGE:
500.0 Conc
0 0 5 0 0 .0
ENTR
EXIT
SETUP X.X
LOW RANGE:
500.0 Conc
0 0 5 0 0 .0
ENTR
EXIT
Toggle the numeral
buttons to set the
LOW
and HIGH
range value.
ENTR accepts the
new setting, EXIT
ignores the new
setting.
Figure 5-6: SETUP RNGE – Auto Range Mode
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
In AUTO range mode, the LOW and HIGH ranges have separate slopes
and offsets for computing SO
2
concentration. The two ranges must be
independently calibrated.
5.4.4. Range Units
The 6200T can display concentrations in parts per billion (10
9
mols per mol, PPB), parts
per million (10
6
mols per mol, PPM), micrograms per cubic meter (µg/m
3
, UGM) or