Teledyne 6200T - Sulfides Analyzer User Manual
Page 164
DAS and AICOM
Model 6200T Total Sulfur Analyzer
Teledyne Analytical Instruments
164
When editing the data channels, the top line of the display indicates some of the
configuration parameters. For example, the display line:
0) CONC: ATIMER, 4, 800
represents to the following configuration:
CHANNEL NUMBER.: 0
NAME: CONC
TRIGGER EVENT: ATIMER
PARAMETERS: Four parameters are included in this channel
EVENT: This channel is set up to record 800 data points.
To edit the name of a data channel, refer to Figure 7-3, then press:
SETUP X.X
NAME:CONC
C O N C 1 -
ENTR
EXIT
ENTR
accepts the new string
and returns to the previous
menu.
EXIT
ignores the new string
and returns to the previous
menu.
Press each as many times as needed to cycle through the
character set until desired character appears:
0-9, A-Z, space ’ ~ !
# $ % ^ & * ( ) - _ = +[ ] { } < >\ | ; : , . / ?
SETUP X.X
NAME:CONC1
<SET SET>
EDIT
EXIT
Figure 7-4: DAS – Editing Data Channel Name