DAS and AICOM

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

164

When editing the data channels, the top line of the display indicates some of the
configuration parameters. For example, the display line:

0) CONC: ATIMER, 4, 800

represents to the following configuration:

CHANNEL NUMBER.: 0
NAME: CONC
TRIGGER EVENT: ATIMER
PARAMETERS: Four parameters are included in this channel
EVENT: This channel is set up to record 800 data points.

To edit the name of a data channel, refer to Figure 7-3, then press:


SETUP X.X

NAME:CONC


C O N C 1 -

ENTR

EXIT

ENTR

accepts the new string

and returns to the previous

menu.

EXIT

ignores the new string

and returns to the previous

menu.

Press each as many times as needed to cycle through the

character set until desired character appears:

0-9, A-Z, space ’ ~ !

# $ % ^ & * ( ) - _ = +[ ] { } < >\ | ; : , . / ?

SETUP X.X

NAME:CONC1


<SET SET>

EDIT

PRINT

EXIT

Figure 7-4: DAS – Editing Data Channel Name