6200T Total Sulfur Analyzer

Calibration Procedures

Teledyne Analytical Instruments

199

STEP TWO: Perform the zero/span check.


Wait until STABIL

falls below 0.5
ppb. This may

take several

minutes.

The value of STABIL

may jump

significantly. Wait
until STABIL falls

below 0.5 ppb. This

may take several

minutes.

SAMPLE

STABIL=XXX.X PPB

SO2 =XXX.X

< TST TST > CAL

CALZ

CALS

SETUP

ZERO CAL M

STABIL=XXX.X PPB

SO2 =XXX.X

< TST TST >

ZERO

CONC

EXIT

SAMPLE

STABIL=XXX.X PPB

SO2 =XXX.X

< TST TST > CAL

CALZ

CALS

SETUP

EXIT

returns to the main

SAMPLE

display

ACTION:

Record the

SO

2

readings

presented in the

upper right corner of

the display.

ACTION:

Record the

SO

2

readings

presented in the

upper right corner of

the display.

Scroll to the STABIL test

function.

SAMPLE RANGE = 500.000 PPB SO2 =XXX.X

< TST TST >

CAL

CALZ

CALS

SETUP

SPAN CAL M

STABIL=XXX.X PPB

SO2 =XXX.X

< TST TST >

ZERO

SPAN

CONC

EXIT

Figure 9-10:

Setup for Manual Calibration Check with Z/S Valve or IZS Option (Step 2)