Teledyne 6200T - Sulfides Analyzer User Manual
Page 344
Instrument Test and Calibration Record
Model 6200T Total Sulfur Analyzer
344
Teledyne Analytical Instruments
Table 17-3.
Test and Calibrations Values w/ CO2 where applicable
Span and Cal
Value
s
Acceptable Value
Parameter
Observed Value
Units
Nominal Range
TS Span Conc.
PPB
20 - 20,000
TS Slope
1.0 +/- .3
TS Offset
MV
< 100
Noise at Zero (rms)
PPB
< 0.2
Noise at Span (rms)
PPB
< 0.5
PMT at Zero
(SO
2
/CO
2
)
MV
PMT at Span
(SO
2
/CO
2
)
MV
Measured Flows
Parameter
Observed Value
Units
Nominal Range
Sample Flow w/ CO
2
cc/min 500
-
700
Sample Flow w/Air
cc/min
400 - 600
Sample Press w/CO
2
“ HG
24 - 27
IZS Purge Flow
cc/min
50 +/- 10
H
2
S Conversion
Efficiency
Expected =
______
PPB
Actual = ______
PPB
Efficiency = ________ %
(100
2%)
Factory Installed Options
Option Installed
Power Voltage/Frequency
Rack Mount, w/ Slides
Rack Mount, w/ Ears Only
Internal Zero/Span - IZS
Permeation Tube (Output Specification)
4-20 MA Current Loop Output
External Pump
PROM Rev #: _______________________
6200TTS S/N: ____________
M501TS S/N: ____________
Date: _______________________ Technician: _______________________