Instrument Test and Calibration Record

Model 6200T Total Sulfur Analyzer

344

Teledyne Analytical Instruments

Table 17-3.

Test and Calibrations Values w/ CO2 where applicable

Span and Cal

Value
s

Acceptable Value

Parameter

Observed Value

Units

Nominal Range

TS Span Conc.

PPB

20 - 20,000

TS Slope

1.0 +/- .3

TS Offset

MV

< 100

Noise at Zero (rms)

PPB

< 0.2

Noise at Span (rms)

PPB

< 0.5

PMT at Zero
(SO

2

/CO

2

)

MV

PMT at Span
(SO

2

/CO

2

)

MV

Measured Flows

Parameter

Observed Value

Units

Nominal Range

Sample Flow w/ CO

2

cc/min 500

-

700

Sample Flow w/Air

cc/min

400 - 600

Sample Press w/CO

2

“ HG

24 - 27

IZS Purge Flow

cc/min

50 +/- 10

H

2

S Conversion

Efficiency

Expected =
______

PPB

Actual = ______
PPB

Efficiency = ________ %

(100

 2%)

Factory Installed Options

Option Installed

Power Voltage/Frequency

Rack Mount, w/ Slides

Rack Mount, w/ Ears Only

Internal Zero/Span - IZS

Permeation Tube (Output Specification)

4-20 MA Current Loop Output

External Pump

PROM Rev #: _______________________

6200TTS S/N: ____________

M501TS S/N: ____________

Date: _______________________ Technician: _______________________