6200t total sulfur analyzer communications setup, Changing the analyzer’s hostname, Teledyne analytical instruments 147 – Teledyne 6200T - Sulfides Analyzer User Manual

Page 147

Advertising
background image

6200T Total Sulfur Analyzer

Communications Setup

Teledyne Analytical Instruments

147

6.5.2. Configuring Ethernet Communication Using Dynamic Host

Configuration Protocol (DHCP)

1. Consult with your network administrator to affirm that your network server is running

DHCP.

2. Access the Communications Menu as shown in Figure 5-13.

3. Follow the setup sequence as shown in Figure 6-5.


SETUP X.X

COMMUNICATIONS MENU

ID

INET

COM1

COM2

EXIT

SETUP X.X

INST IP: 0.0.0.0

<SET

SET>

EXIT

SETUP X.X

GATEWAY IP: 0.0.0.0

<SET

SET>

EXIT

SETUP X.X

TCP PORT2: 502

<SET

SET>

EDIT

EXIT

EDIT

button

disabled

From this point on,

EXIT

returns to

COMMUNICATIONS

MENU

Do not alter unless

directed to by Teledyne

Instruments Customer

Service personnel

SETUP X.X

HOSTNAME:

<SET

EDIT

EXIT

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

SUBNET MASK: 0.0.0.0

<SET

SET>

EXIT

SETUP X.X TCP PORT: 3000

<SET

SET>

EDIT

EXIT

SETUP X.X

DHCP: ON

SET>

EDIT

EXIT

DHCP: ON is

default setting.

If it has been

set to OFF,
press EDIT

and set to ON.

SETUP X.X

DHCP: OFF

OFF

ENTR

EXIT

SETUP X.X

DHCP: ON

ON

ENTR

EXIT

Figure 6-5: COMM – LAN / Internet Automatic Configuration

6.5.2.1. Changing the Analyzer’s HOSTNAME

The HOSTNAME is the name by which the analyzer appears on your network. The
default name for all Teledyne Analytical Instruments’s 6200T analyzers is 6200T. To