6200t total sulfur analyzer communications setup, Changing the analyzer’s hostname, Teledyne analytical instruments 147 – Teledyne 6200T - Sulfides Analyzer User Manual
Page 147
6200T Total Sulfur Analyzer
Communications Setup
Teledyne Analytical Instruments
147
6.5.2. Configuring Ethernet Communication Using Dynamic Host
Configuration Protocol (DHCP)
1. Consult with your network administrator to affirm that your network server is running
DHCP.
2. Access the Communications Menu as shown in Figure 5-13.
3. Follow the setup sequence as shown in Figure 6-5.
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
COM2
EXIT
SETUP X.X
INST IP: 0.0.0.0
<SET
SET>
EXIT
SETUP X.X
GATEWAY IP: 0.0.0.0
<SET
SET>
EXIT
SETUP X.X
TCP PORT2: 502
<SET
SET>
EDIT
EXIT
EDIT
button
disabled
From this point on,
EXIT
returns to
COMMUNICATIONS
MENU
Do not alter unless
directed to by Teledyne
Instruments Customer
Service personnel
SETUP X.X
HOSTNAME:
<SET
EDIT
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
SUBNET MASK: 0.0.0.0
<SET
SET>
EXIT
SETUP X.X TCP PORT: 3000
<SET
SET>
EDIT
EXIT
SETUP X.X
DHCP: ON
SET>
EDIT
EXIT
DHCP: ON is
default setting.
If it has been
set to OFF,
press EDIT
and set to ON.
SETUP X.X
DHCP: OFF
OFF
ENTR
EXIT
SETUP X.X
DHCP: ON
ON
ENTR
EXIT
Figure 6-5: COMM – LAN / Internet Automatic Configuration
6.5.2.1. Changing the Analyzer’s HOSTNAME
The HOSTNAME is the name by which the analyzer appears on your network. The
default name for all Teledyne Analytical Instruments’s 6200T analyzers is 6200T. To