6200t total sulfur analyzer setup menu – Teledyne 6200T - Sulfides Analyzer User Manual

Page 123

Advertising
background image

6200T Total Sulfur Analyzer

Setup Menu

Teledyne Analytical Instruments

123

To automatically calibrate a single analog channel from the DIAG Menu (refer to Figure
5-16), press:

DIAG AIO

CONC_OUT_2 CALIBRATED: NO

<SET

CAL

EXIT

DIAG AIO

AUTO CALIBRATING CONC_OUT_2

DIAG AIO

CONC_OUT_2 CALIBRATED:

YES

<SET CAL

EXIT

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

DIAG AIO

AOUTS CALIBRATED:

NO

<

SET>

CAL

EXIT

DIAG AIO

CONC_OUT_2 RANGE: 5V

SET>

EDIT

EXIT

DIAG AIO

CONC_OUT_2:5V, CAL

< SET SET>

EDIT

EXIT

DIAG AIO

CONC_OUT_2 REC OFS: 0 mV

< SET

SET>

EDIT

EXIT

DIAG AIO

CONC_OUT_2 AUTO CAL: ON

< SET

SET> EDIT

EXIT

EXIT

to Return

to the main

Sample Display

Press SET> to select the
Analog Output channel to
be configured. Then Press
EDIT

to continue

Figure 5-21:

DIAG – Analog Output Calibration Mode – Single Analog Channel