background image

6200T Total Sulfur Analyzer

Communications Setup

Teledyne Analytical Instruments

153

6.6.2.3. Selecting a Hessen Protocol Type

Currently there are two version of Hessen Protocol in use. The original implementation,
referred to as TYPE 1, and a more recently released version, TYPE 2 that has more
flexibility when operating with instruments that can measure more than one type of gas.
For more specific information about the difference between TYPE 1 and TYPE 2
download the Manual Addendum for Hessen Protocol from Teledyne’s web site:

http://www.teledyne-api.com/manuals/index.asp.

To select a Hessen Protocol Type, access the COMMUNICATIONS Menu (refer to
Figure 5-13), then press:

SETUP X.X

COMMUNICATIONS MENU

ID INET

HESN

COM1

COM2 EXIT

SETUP X.

HESSEN VARIATION: TYPE 1

SET>

EDIT

EXIT

SETUP X.X

HESSEN VARIATION: TYPE 1

TYPE1 TYPE 2

ENTR EXIT

SETUP X.X

HESSEN VARIATION: TYPE 2

PREV NEXT

OFF

ENTR EXIT

Press to change

protocol type.

ENTR accepts the new

settings

EXIT ignores the new

settings

Figure 6-8: COMM – Select Hessen Protocol Type

Note

While Hessen Protocol Mode can be activated independently for RS-232
and COM2, the TYPE selection affects both Ports.