6200t total sulfur analyzer communications setup, Selecting a hessen protocol type, Teledyne analytical instruments 153 – Teledyne 6200T - Sulfides Analyzer User Manual
Page 153: Figure 6-8: comm – select hessen protocol type
6200T Total Sulfur Analyzer
Communications Setup
Teledyne Analytical Instruments
153
6.6.2.3. Selecting a Hessen Protocol Type
Currently there are two version of Hessen Protocol in use. The original implementation,
referred to as TYPE 1, and a more recently released version, TYPE 2 that has more
flexibility when operating with instruments that can measure more than one type of gas.
For more specific information about the difference between TYPE 1 and TYPE 2
download the Manual Addendum for Hessen Protocol from Teledyne’s web site:
http://www.teledyne-api.com/manuals/index.asp.
To select a Hessen Protocol Type, access the COMMUNICATIONS Menu (refer to
Figure 5-13), then press:
SETUP X.X
COMMUNICATIONS MENU
ID INET
HESN
COM1
COM2 EXIT
SETUP X.
HESSEN VARIATION: TYPE 1
SET>
EDIT
EXIT
SETUP X.X
HESSEN VARIATION: TYPE 1
TYPE1 TYPE 2
ENTR EXIT
SETUP X.X
HESSEN VARIATION: TYPE 2
PREV NEXT
OFF
ENTR EXIT
Press to change
protocol type.
ENTR accepts the new
settings
EXIT ignores the new
settings
Figure 6-8: COMM – Select Hessen Protocol Type
Note
While Hessen Protocol Mode can be activated independently for RS-232
and COM2, the TYPE selection affects both Ports.