Figure 5-5: setup rnge – dual range mode – Teledyne 6200T - Sulfides Analyzer User Manual

Page 102

Advertising
background image

Setup Menu

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

102

5.4.3.2. Dual Range Mode (DUAL)

Selecting Dual Range mode allows the A1 and A2 outputs to be configured with
different reporting ranges. The analyzer software calls these two ranges low and high.
The Low range setting corresponds with the analog output labeled A1 on the rear panel
of the instrument. The high range setting corresponds with the A2 output. While the
software names these two ranges low and high, they do not have to be configured that
way. For example: the low range can be set for a span of 0-150 ppb while the high range
is set for 0-50 ppb.
In DUAL range mode the RANGE test function displayed on the front panel will be
replaced by two separate functions:

 Range1: The range setting for the A1 output.
 Range2: The range setting for the A2 output.

To set the ranges press following control button sequence

.

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK MORE

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

SETUP X.X RANGE

MODE:

SNGL

SNGL

DUAL

AUTO

ENTR EXIT

SETUP X.X

RANGE MODE:

DUAL

SNGL

DUAL AUTO

ENTR

EXIT

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET

UNIT

EXIT

SETUP X.X

HIGH RANGE: 500.0 Conc

0 0 5

0 0 .0

ENTR

EXIT

SETUP X.X

LOW RANGE

:

500.0 Conc

0 0 1 0 0 .0

ENTR

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

EXIT

Returns

to the Main

SAMPLE

Display

Toggle the

numeral buttons

to set the upper

limit of each

range.

SAMPLE RANGE = 500.000 PPB SO2 =XXX.X

< TST TST > CAL

SETUP

Figure 5-5: SETUP RNGE – Dual Range Mode

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

In DUAL range mode the LOW and HIGH ranges have separate slopes
and offsets for computing SO

2

concentration. The two ranges must be

independently calibrated.