6200t total sulfur analyzer setup menu, Teledyne analytical instruments 129 – Teledyne 6200T - Sulfides Analyzer User Manual

Page 129

Advertising
background image

6200T Total Sulfur Analyzer

Setup Menu

Teledyne Analytical Instruments

129

To adjust the zero and span values of the current outputs, activate the ANALOG I/O
CONFIGURATION
MENU from the DIAG Menu (refer to Figure 5-16), then press:


EXAMPLE

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

DIAG AIO

AIN A/C FREQUENCY:

60 HZ


SET>

EDIT

EXIT

DIAG AIO

AOUT CALIBRATED:

NO


< SET

SET>

CAL

EXIT

DIAG AIO

CONC_OUT_2 RANGE: CURR


<SET

SET>

EDIT

EXIT

DIAG AIO

CONC_OUT_2 CALIBRATED: NO

< SET

CAL

EXIT

DIAG AIO

CONC_OUT_2 ZERO: 0 mV


U100

UP10

UP DOWN DN10 D100 ENTR EXIT

DIAG AIO

CONC_OUT_CURR, NO CAL

< SET SET>

EDIT

EXIT

Press SET> to select the analog output channel

to be configured:. Then press EDIT to continue

DIAG AIO

AIN CALIBRATED:

NO


SET>

EDIT

EXIT

ENTR r

eturns

to the previous

menu.

Increase or decrease the current

output by 100, 10 or 1 counts. The

resulting change in output voltage is

displayed in the upper line.

Continue adjustments until the correct

current is measured with the current

meter.

DIAG AIO

CONC_OUT_2 ZERO: 27 mV


U100 UP10

UP DOWN DN10 D100

ENTR EXIT

EXIT

ignores the

new setting, ENTR

accepts the new

setting.

DIAG AIO

CONC_OUT_2 CALIBRATED: YES

< SET

CAL

EXIT

DIAG AIO

CONC_OUT_2 SPAN: 10000 mV


U100 UP10

UP DOWN DN10

D100

ENTR EXIT

EXAMPLE

DIAG AIO

CONC_OUT_2 ZERO: 9731 mV


U100 UP10

UP DOWN DN10 D100

ENTR EXIT

Figure 5-27:

Analog Output – Zero and Span Value Adjustment for Current Outputs

If a current meter is not available, an alternative method for calibrating the current loop outputs is to connect a
250

 1% resistor across the current loop output. Using a voltmeter, connected across the resistor, follow the

procedure above but adjust the output to the following values:

Table 5-8: Current Loop Output Calibration with Resistor

FULL SCALE

VOLTAGE FOR 2-20 MA

(MEASURED ACROSS

RESISTOR)

VOLTAGE FOR 4-20 MA

(MEASURED ACROSS RESISTOR)

0%

0.5 V

1.0 V

100%

5.0 V

5.0 V