To adjust the zero and span values of the current outputs, activate the ANALOG I/O
CONFIGURATION MENU from the DIAG Menu (refer to Figure 5-16), then press:
EXAMPLE
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AIN A/C FREQUENCY:
60 HZ
SET>
EDIT
EXIT
DIAG AIO
AOUT CALIBRATED:
NO
< SET
SET>
CAL
EXIT
DIAG AIO
CONC_OUT_2 RANGE: CURR
<SET
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 CALIBRATED: NO
< SET
CAL
EXIT
DIAG AIO
CONC_OUT_2 ZERO: 0 mV
U100
UP10
UP DOWN DN10 D100 ENTR EXIT
DIAG AIO
CONC_OUT_CURR, NO CAL
< SET SET>
EDIT
EXIT
Press SET> to select the analog output channel
to be configured:. Then press EDIT to continue
DIAG AIO
AIN CALIBRATED:
NO
SET>
EDIT
EXIT
ENTR r
eturns
to the previous
menu.
Increase or decrease the current
output by 100, 10 or 1 counts. The
resulting change in output voltage is
displayed in the upper line.
Continue adjustments until the correct
current is measured with the current
meter.
DIAG AIO
CONC_OUT_2 ZERO: 27 mV
U100 UP10
UP DOWN DN10 D100
ENTR EXIT
EXIT
ignores the
new setting, ENTR
accepts the new
setting.
DIAG AIO
CONC_OUT_2 CALIBRATED: YES
< SET
CAL
EXIT
DIAG AIO
CONC_OUT_2 SPAN: 10000 mV
U100 UP10
UP DOWN DN10
D100
ENTR EXIT
EXAMPLE
DIAG AIO
CONC_OUT_2 ZERO: 9731 mV
U100 UP10
UP DOWN DN10 D100
ENTR EXIT
Figure 5-27:
Analog Output – Zero and Span Value Adjustment for Current Outputs
If a current meter is not available, an alternative method for calibrating the current loop outputs is to connect a
250
1% resistor across the current loop output. Using a voltmeter, connected across the resistor, follow the
procedure above but adjust the output to the following values:
Table 5-8: Current Loop Output Calibration with Resistor
FULL SCALE
VOLTAGE FOR 2-20 MA
(MEASURED ACROSS
RESISTOR)
VOLTAGE FOR 4-20 MA
(MEASURED ACROSS RESISTOR)
0%
0.5 V
1.0 V
100%
5.0 V
5.0 V