Sensor module repair & cleaning – Teledyne 6200T - Sulfides Analyzer User Manual

Page 262

Advertising
background image

Troubleshooting

Model 6200T Total Sulfur Analyzer

262

Teledyne Analytical Instruments

12.7.2. Sensor Module Repair & Cleaning

CAUTION - GENERAL SAFETY HAZARD

Do not look at the UV lamp while the unit is operating. UV light can cause
eye damage. Always use safety glasses made from UV blocking material
when working with the UV Lamp Assembly. (Generic plastic glasses are
not adequate).

Figure 12-6:

Sensor Module Wiring and Pneumatic Fittings

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

After any repair or service has been performed on the sensor module, the
6200T should be allowed to warm up for 60 minutes.

Always perform a leak check (refer to Section 11.3.6) and calibrate the
analyzer (refer to Section 9) before placing it back in service.