Sensor module repair & cleaning – Teledyne 6200T - Sulfides Analyzer User Manual
Page 262
Advertising
Troubleshooting
Model 6200T Total Sulfur Analyzer
262
Teledyne Analytical Instruments
12.7.2. Sensor Module Repair & Cleaning
CAUTION - GENERAL SAFETY HAZARD
Do not look at the UV lamp while the unit is operating. UV light can cause
eye damage. Always use safety glasses made from UV blocking material
when working with the UV Lamp Assembly. (Generic plastic glasses are
not adequate).
Figure 12-6:
Sensor Module Wiring and Pneumatic Fittings
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
After any repair or service has been performed on the sensor module, the
6200T should be allowed to warm up for 60 minutes.
Always perform a leak check (refer to Section 11.3.6) and calibrate the
analyzer (refer to Section 9) before placing it back in service.