Manual calibration checks, 6200t total sulfur analyzer calibration procedures, Teledyne analytical instruments 193 – Teledyne 6200T - Sulfides Analyzer User Manual

Page 193: Figure 9-4: setup for manual calibration checks

Advertising
background image

6200T Total Sulfur Analyzer

Calibration Procedures

Teledyne Analytical Instruments

193

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

If the ZERO or SPAN buttons are not displayed during zero or span
calibration, the measured concentration value is too different from the
expected value and the analyzer does not allow zeroing or spanning the
instrument. Refer to Section 12.4 for more information on calibration
problems.

9.3. Manual Calibration Checks

Informal calibration checks will only evaluate the analyzer’s response curve, but do not
alter it. It is recommended as a regular maintenance item, to perform calibration checks
in order to monitor the analyzer’s performance. To carry out a calibration check rather
than a full calibration, perform the following procedures:

STEP ONE: Connect the sources of zero air and span gas as shown in
Figure 9-1.

STEP TWO: Perform the zero/span calibration check procedure:

SAMPLE

STABIL=XXX.X PPB

SO2=XXX.X


< TST TST >

CAL

SETUP

ACTION:

Record the SO

2

concentration

reading.

ACTION:

Supply span gas to the instrument

SAMPLE

STABIL=XXX.X PPB

SO2=XXX.X


< TST TST >

CAL

SETUP

ACTION:

Record the SO

2

concentration

reading.

Wait until

STABIL

is

below 0.5 ppb.

This may take

several minutes.

The value of

STABIL

may jump

significantly.

Wait until it falls

below 0.5 ppb. This

may take several

minutes.

SAMPLE

STABIL=XXX.X PPB

SO2=XXX.X

< TST TST >

CAL

SETUP

The SPAN button appears during the transition from zero to

span. You may see both the SPAN and the ZERO buttons.

SAMPLE RANGE = 500.000 PPB SO2 =XXX.X

< TST TST >

CAL

SETUP

Scroll the display to the

STABIL

test function.

ACTION:

Supply the instrument with zero gas.

Figure 9-4: Setup for Manual Calibration Checks

Advertising