Communications protocols, Modbus, 6200t total sulfur analyzer communications setup – Teledyne 6200T - Sulfides Analyzer User Manual

Page 149: Modbus setup

Advertising
background image

6200T Total Sulfur Analyzer

Communications Setup

Teledyne Analytical Instruments

149

1. Go to SETUP>MORE>COMM>COM2 menu.
2. Press the SET> button until “COM2 BAUD RATE:xxxxx” appears in the Param

field of the instrument display. (Figure 6-2 shows the menu sequence).

3. Check that the baud rate of the instrument matches the baud rate of your PC (if

needed change either one to match the other).

4. Press the ENTR button to accept any changes.

6.6. Communications Protocols

6.6.1. MODBUS

The following set of instructions assumes that the user is familiar with MODBUS
communications, and provides minimal information to get started. For additional
instruction, please refer to the Teledyne Analytical Instruments MODBUS manual, PN
06276. Also refer to www.modbus.org for MODBUS communication protocols.

Minimum Requirements

Instrument firmware with MODBUS capabilities installed.

MODBUS-compatible software (TAI uses MODBUS Poll for testing; see
www.modbustools.com)

Personal computer

Communications cable (Ethernet or USB or RS232)

Possibly a null modem adapter or cable

MODBUS Setup:

Set Com Mode parameters

Comm







Slave ID

Ethernet: Using the front panel menu, go to SETUP – MORE – COMM – INET; scroll through the INET

submenu until you reach TCP PORT 2 (the standard setting is 502), then continue to TCP
PORT 2 MODBUS TCP/IP; press EDIT and toggle the menu button to change the setting
to ON, then press ENTR. (Change Machine ID if needed: see “Slave ID”).

USB/RS232: Using the front panel menu, go to SETUP – MORE – COMM – COM2 – EDIT; scroll through

the COM2 EDIT submenu until the display shows COM2 MODBUS RTU: OFF (press
OFF to change the setting to ON. Scroll NEXT to COM2 MODBUS ASCII and ensure it is
set to OFF. Press ENTR to keep the new settings. (If RTU is not available with your
communications equipment, set the COM2 MODBUS ASCII setting to ON and ensure that
COM2 MODBUS RTU is set to OFF. Press ENTR to keep the new settings).


A MODBUS slave ID must be set for each instrument. Valid slave ID’s are in the range of 1 to 247. If
your analyzer is connected to a serial network (i.e., RS-485), a unique Slave ID must be assigned to each
instrument. To set the slave ID for the instrument, go to SETUP – MORE – COMM – ID. The default
MACHINE ID is the same as the model number. Toggle the menu buttons to change the ID.

Reboot analyzer

For the settings to take effect, power down the analyzer, wait 5 seconds, and power up the analyzer.

Make appropriate cable
connections

Connect your analyzer either:

via its Ethernet or USB port to a PC (this may require a USB-to-RS232 adapter for your PC; if so, also
install the software driver from the CD supplied with the adapter, and reboot the computer if required), or

via its COM2 port to a null modem (this may require a null modem adapter or cable).

Specify MODBUS software
settings
(examples used here are for
MODBUS Poll software)

1.

Click Setup / [Read / Write Definition] /.

a.

In the Read/Write Definition window (see example that follows) select a Function (what you wish
to read from the analyzer).

b.

Input Quantity (based on your firmware’s register map).

c.

In the View section of the Read/Write Definition window select a Display (typically Float Inverse).

d.

Click OK.

2.

Next, click Connection/Connect.

a.

In the Connection Setup window (see example that follows), select the options based on your
computer.

b.

Press OK.

Read the Modbus Poll Register Use the Register Map to find the test parameter names for the values displayed (see example that follows

If desired, assign an alias for each.