Warning messages – Teledyne 6200T - Sulfides Analyzer User Manual

Page 93

Advertising
background image

6200T Total Sulfur Analyzer

Overview of Operating Modes

Teledyne Analytical Instruments

93

4.1.2. Warning Messages

The most common instrument failures will be reported as a warning on the analyzer’s
front panel and through the COMM ports. Section 12.1.1 explains how to use these
messages to troubleshoot problems. Section 12.1.3 shows how to view and clear
warning messages. Table 4-3 lists all warning messages for the current version of
software.

Table 4-3: List of Warning Messages

MESSAGE MEANING

ANALOG CAL WARNING

The instrument's A/D circuitry or one of its analog outputs is not calibrated.

BOX TEMP WARNING

The temperature inside the 6200T chassis is outside the specified limits.

CANNOT DYN SPAN

Remote span calibration failed while the dynamic span feature was set to turned on

CANNOT DYN ZERO

Remote zero calibration failed while the dynamic zero feature was set to turned on

CONFIG INITIALIZED

Configuration was reset to factory defaults or was erased.

DARK CAL WARNING

Dark offset above limit specified indicating that too much stray light is present in the
sample chamber.

DATA INITIALIZED

DAS data storage was erased.

HVPS WARNING

High voltage power supply for the PMT is outside of specified limits.

IZS TEMP WARNING

On units with IZS options installed: The permeation tube temperature is outside of
specified limits.

PMT DET WARNING

PMT detector output outside of operational limits.

PMT TEMP WARNING

PMT temperature is outside of specified limits.

RCELL TEMP WARNING

Sample chamber temperature is outside of specified limits.

REAR BOARD NOT DET

The CPU is unable to communicate with the motherboard.

RELAY BOARD WARN

The firmware is unable to communicate with the relay board.

SAMPLE FLOW WARN

The flow rate of the sample gas is outside the specified limits.

SAMPLE PRESS WARN

Sample pressure outside of operational parameters.

SYSTEM RESET

The computer was rebooted.

UV LAMP WARNING

The UV lamp intensity measured by the reference detector reading too low or too high

UUUU

(M501TS Converter) Open thermocouple