Teledyne 6200T - Sulfides Analyzer User Manual
Page 76
Getting Started
Model 6200T Total Sulfur Analyzer
Teledyne Analytical Instruments
76
Table includes typical values for a converter set up for operation on 115V/60Hz. The
P, I and d values may be different for individual converters and AC mains voltages, and
will vary somewhat after auto-tuning.
Table 3-13. Temperature Controller – Initial Settings
PRESS DISPLAY
INITIAL
VALUE
SEL
P
PROP BAND
UP/DOWN SET TO “11”
SEL
I
INTEGRAL
SET TO “10”
SEL
d
DERIVATIVE TIME
SET TO “7.7”
SEL
AL
LOW ALARM SETPOINT
SET TO “50” (ºC BELOW FINAL SETVALUE)
SEL
AH
HIGH ALARM SETPOINT
SET TO “50” (ºC ABOVE FINAL SETVALUE)
SEL
7C
CYCLE TIME
SET TO “2”
SEL
HYS
HYSTERESIS
SET TO “3”
SEL
A7
AUTOTUNE
SET TO “0” (OFF)
SEL
LOC
LOCK
“0” (OPEN) “1” (LOCKED)
“2” (SV ONLY OPEN)
Table shows a Secondary Menu of parameters that set more basic parameters of the
controller, these include the thermocouple type, the temperature units etc.
CAUTION – Avoid Damage to the Instrument
Use only the “p-n2” setting specified for the specific thermocouple Type as
described in Table for “p n2”. If in doubt about which thermocouple Type you have,
please contact Teledyen Analytical InstrumentsCustomer Service.
Table 3-14. Temperature Controller - Secondary Menu
PRESS
DISPLAY
3.4.1.3.0.0.0.1 PARAMETER VALUE
SEL
HOLD TILL p-n1
SET TO “0”
SEL
p-n2*
* Set only to the value
stated in the next
column.
Type S thermocouple (red & black wires): SET TO “6”
In the rare event that you have a different
thermocouple:
Type K, red and yellow wires, SET TO “3”_
Type N, red and orange wires, SET TO “12”.
CAREFULLY FOLLOW THE INSTALLATION
INSTRUCTIONS PROVIDED IN THE
THERMOCOUPLE REPLACEMENT SECTION OF
THIS MANUAL.