Getting Started

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

76

Table includes typical values for a converter set up for operation on 115V/60Hz. The
P, I and d values may be different for individual converters and AC mains voltages, and
will vary somewhat after auto-tuning.

Table 3-13. Temperature Controller – Initial Settings

PRESS DISPLAY

INITIAL

VALUE

SEL

P

PROP BAND

UP/DOWN SET TO “11”

SEL

I

INTEGRAL

SET TO “10”

SEL

d

DERIVATIVE TIME

SET TO “7.7”

SEL

AL

LOW ALARM SETPOINT

SET TO “50” (ºC BELOW FINAL SETVALUE)

SEL

AH

HIGH ALARM SETPOINT

SET TO “50” (ºC ABOVE FINAL SETVALUE)

SEL

7C

CYCLE TIME

SET TO “2”

SEL

HYS

HYSTERESIS

SET TO “3”

SEL

A7

AUTOTUNE

SET TO “0” (OFF)

SEL

LOC

LOCK

“0” (OPEN) “1” (LOCKED)

“2” (SV ONLY OPEN)


Table shows a Secondary Menu of parameters that set more basic parameters of the
controller, these include the thermocouple type, the temperature units etc.

CAUTION – Avoid Damage to the Instrument

Use only the “p-n2” setting specified for the specific thermocouple Type as
described in Table for “p n2”. If in doubt about which thermocouple Type you have,
please contact Teledyen Analytical InstrumentsCustomer Service.

Table 3-14. Temperature Controller - Secondary Menu

PRESS

DISPLAY

3.4.1.3.0.0.0.1 PARAMETER VALUE

SEL

HOLD TILL p-n1

SET TO “0”

SEL

p-n2*

* Set only to the value

stated in the next
column.

Type S thermocouple (red & black wires): SET TO “6”
In the rare event that you have a different
thermocouple:
Type K, red and yellow wires, SET TO “3”_
Type N, red and orange wires, SET TO “12”.

CAREFULLY FOLLOW THE INSTALLATION
INSTRUCTIONS PROVIDED IN THE
THERMOCOUPLE REPLACEMENT SECTION OF
THIS MANUAL.