Table 3-14:

Possible Startup Warning Messages – 6200T Analyzers with Options

Message Meaning

O2 CELL TEMP WARN

1

O

2

sensor cell temperature outside of warning limits specified by

O2_CELL_SET

variable.

IZS TEMP WARNING

2

On units with IZS options installed: The permeation tube temperature is outside
of specified limits.

O2 ALARM 1 WARN

1, 4

O

2

Alarm limit #1 has been triggered.

4

O2 ALARM 2 WARN

1, 4

O

2

Alarm limit #2 has been triggered.

4

CO2 ALARM 1 WARN

3, 4

CO

2

Alarm limit #1 has been triggered.

4

CO2 ALARM 2 WARN

3, 4

`

CO

2

Alarm limit #2 has been triggered.

4

SO2 ALARM1 WARN

4

SO

2

Alarm limit #1 has been triggered.

4

SO2 ALARM2 WARN

4

SO

2

Alarm limit #2 has been triggered.

4

1

Only appears when the optional O

2

sensor is installed.

2

Only appears when the optional internal zero span (IZS) option is installed.

3

Only appears when the optional CO

2

sensor is installed.

4

Only Appears when the optional gas concentration alarms are installed

3.4.3. Functional Checks

After the analyzer’s components have warmed up for at least 60 minutes and the
temperature inside the quartz tube has stabilized at the setpoint, verify that the software
properly supports any hardware options that were installed.
For information on navigating through the analyzer’s software menus, refer to the menu
trees described in Appendix A.1.
Check to ensure that the analyzer is functioning within allowable operating parameters.

 Appendix C includes a list of test functions viewable from the analyzer’s

front panel as well as their expected values.

 These functions are also useful tools for diagnosing performance problems

with your analyzer (refer to Section 12.1.2).

 The enclosed Final Test and Validation Data Sheet (P/N 04551) lists these

values before the instrument left the factory.