Calibration Procedures

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

192

M-P CAL

SO2 STB=X.XXX PPB

SO2

=XXX.X

< TST TST >

ENTR

CONC

EXIT

M-P CAL

SO2 STB=X.XXX PPB

SO2

=XXX.X

< TST TST >

CAL

SETUP

M-P CAL

SO2 STB=X.XXX PPB

SO2

=XXX.X

< TST TST >

ZERO

CONC

EXIT

EXIT

returns to the main

SAMPLE

display

ACTION:

Allow zero gas to enter the sample port at the

rear of the instrument.

M-P CAL

SO2 STB=X.XXX PPB

SO2

=XXX.X

< TST TST >

SPAN

CONC

EXIT

M-P CAL RANGE = 500.0 PPB

SO2

=XXX.X

< TST TST >

ENTR

SPAN CONC

EXIT

Wait until SO2 STB

falls below 0.5 ppb.

This may take several

minutes.

The SPAN button now

appears during the

transition from zero to span.

You may see both the

SPAN and the ZERO

buttons.

SAMPLE RANGE = 500.0 PPB SO2 =XXX.X

< TST TST >

CAL

SETUP

Set the Display to show the

SO2 STB

test function.

This function calculates the

stability of the SO

2

measurement

Press ENTR to changes the

OFFSET

& SLOPE values for the

SO

2

measurements.

Press EXIT to leave the calibration

unchanged and return to the

previous menu.

ACTION:

Allow span gas to enter the sample port at the

rear of the instrument.

The value of

SO2 STB

may jump

significantly.

Wait until it falls back

below 0.5 ppb.

This may take several

minutes.

Press ENTR to change the

OFFSET

& SLOPE values for the

SO

2

measurements.

Press EXIT to leave the calibration

unchanged and return to the

previous menu.

SAMPLE RANGE = 500.0 PPB SO2 =XXX.X

< TST

TST

>

CAL

SETUP

M-P CAL

RANGE = 500.0 PPB

SO2

=XXX.X

< TST TST > ENTR CONC

EXIT

Figure 9-3: Setup for Manual Calibration without Z/S valve or IZS Option (Step 3)