6200T Total Sulfur Analyzer

Communications Setup

Teledyne Analytical Instruments

155

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

It is possible to assign more than one flag to the same Hessen status bit.
This allows the grouping of similar flags, such as all temperature
warnings, under the same status bit. Be careful not to assign conflicting
flags to the same bit as each status bit will be triggered if any of the
assigned flags are active.

PMT DET WARNING

0002

UV LAMP WARNING

0002

HVPS WARNING

0004

DARK CAL WARNING

0008

RCELL TEMP WARNING

0010

IZS TEMP WARNING

0020

PMT TEMP WARNING

0040

INVALID CONC

0080

OPERATIONAL FLAGS

In Manual Calibration Mode

0200

In Zero Calibration Mode

0400

In Span Calibration Mode

0800

UNITS OF MEASURE FLAGS

UGM 0000

MGM 2000

PPB 4000

PPM 6000

SPARE/UNUSED BITS

100. 8000

UNASSIGNED FLAGS

Box Temp Warning

MP Calibration

Sample Press Warning

Analog Cal Warning

System Reset

Cannot Dyn Zero

Rear Board Not Detected

Cannot Dyn Span

Relay Board Warning

Instrument Off