6200t total sulfur analyzer calibration procedures, Span gas, Step two: set the expected so – Teledyne 6200T - Sulfides Analyzer User Manual
Page 195: Value
6200T Total Sulfur Analyzer
Calibration Procedures
Teledyne Analytical Instruments
195
STEP TWO: Set the expected SO
2
span gas
value:
This sequence causes the
analyzer to prompt for the
expected SO
2
span
concentration.
M-P CAL
SO2 SPAN CONC: 450.0 Conc
0 0 0
4 5 .0
ENTR
EXIT
M-P CAL RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > ZERO
CONC
EXIT
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST >
CAL
SETUP
The SO
2
span concentration
values automatically default
to 450.0 Conc.
To change this value to the
actual concentration of the
span gas, enter the number
by pressing each digit until
the expected value appears.
The span gas concentration
should always be 90% of the
selected reporting range
EXAMPLE
Reporting range = 800 ppb
Span gas conc.= 720 ppb
EXIT
ignores the new setting
and returns to the previous
display.
ENTR
accepts the new setting
and returns to the
previous display..
Figure 9-6: Setup for Manual Calibration with Z/S Valve Option Installed (Step 2)