Setup Menu
Model 6200T Total Sulfur Analyzer
Teledyne Analytical Instruments
124
To select manual output calibration for a particular channel, access the Analog I/O
Configuration from the DIAG Menu (refer to Figure 5-16), then press:
DIAG AIO CONC_OUT_2 AUTO CAL: ON
ON
ENTR EXIT
Exit to return to
the main
sample display
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED: NO
< SET
SET>
CAL
EXIT
DIAG AIO
CONC_OUT_2 RANGE: 5V
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2:5V, CAL
< SET SET>
EDIT
EXIT
Press SET> to select the analog output channel to
be configured. Then press EDIT to continue
ENTR
accepts the new setting
and returns to the previous menu.
EXIT
ignores the new setting and
returns to the previous menu.
DIAG AIO
CONC_OUT_2 REC OFS: 0 mV
< SET
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 AUTO CAL: ON
< SET
SET> EDIT
EXIT
DIAG AIO CONC_OUT_2 AUTO CAL: OFF
OFF ENTR EXIT
Toggle to OFF to enable manual
cal adjustments to the selected
output channel only.
Figure 5-22:
DIAG – Analog Output – Auto Cal or Manual Cal Selection for Channels
Now the analog output channels should either be automatically calibrated or they should
be set to manual calibration, which is described next.