Teledyne 6200T - Sulfides Analyzer User Manual

Page 124

Advertising
background image

Setup Menu

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

124

To select manual output calibration for a particular channel, access the Analog I/O
Configuration from the DIAG Menu (refer to Figure 5-16), then press:

DIAG AIO CONC_OUT_2 AUTO CAL: ON

ON

ENTR EXIT

Exit to return to

the main

sample display

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

DIAG AIO

AOUTS CALIBRATED: NO

< SET

SET>

CAL

EXIT

DIAG AIO

CONC_OUT_2 RANGE: 5V

SET>

EDIT

EXIT

DIAG AIO

CONC_OUT_2:5V, CAL

< SET SET>

EDIT

EXIT

Press SET> to select the analog output channel to

be configured. Then press EDIT to continue

ENTR

accepts the new setting

and returns to the previous menu.

EXIT

ignores the new setting and

returns to the previous menu.

DIAG AIO

CONC_OUT_2 REC OFS: 0 mV

< SET

SET>

EDIT

EXIT

DIAG AIO

CONC_OUT_2 AUTO CAL: ON

< SET

SET> EDIT

EXIT

DIAG AIO CONC_OUT_2 AUTO CAL: OFF

OFF ENTR EXIT

Toggle to OFF to enable manual

cal adjustments to the selected

output channel only.

Figure 5-22:

DIAG – Analog Output – Auto Cal or Manual Cal Selection for Channels

Now the analog output channels should either be automatically calibrated or they should
be set to manual calibration, which is described next.