Teledyne 6200T - Sulfides Analyzer User Manual

Page 63

Advertising
background image

6200T Total Sulfur Analyzer

Getting Started

Teledyne Analytical Instruments

63

Dust Plugs:
Remove dust plugs from rear panel exhaust and supply line fittings

before powering on/operating instrument. These plugs should be kept
for reuse in the event of future storage or shipping to prevent debris
from entering the pneumatics.

I

MPORTANT

EPA Requirements:

US EPA requirements state that zero air and span gases must be

supplied at twice the instrument’s specified gas flow rate. Therefore,
the 6200T zero and span gases should be supplied to their respective
inlets in excess of 1300 cc

3

/min (650 cc

3

/min. x 2).

I

MPORTANT

Leak Check:

Run a leak check once the appropriate pneumatic connections have

been made; check all pneumatic fittings for leaks using the
procedures defined in Section 11.3.6.

CAUTION – GENERAL SAFETY HAZARD

Gas flow though the analyzer must be maintained at all time for units with
a permeation tube installed. Insufficient gas flow allows gas to build up
to levels that will contaminate the instrument or present a safety hazard
to personnel.

Section 3.3.2.1 provides external pneumatic connection instructions, and Table 3-9
provides links to the location of various internal pneumatic layout illustrations.

Table 3-9: Pneumatic Layout Reference

Pneumatic Layout

Section

Basic 3.3.2.2
Zero/Span Valves

3.3.2.3

Internal Zero/Span (IZS)

3.3.2.4

Basic with O2 Sensor

3.3.2.8

Basic with CO2 Sensor

3.3.2.9

Advertising