Gas flow problems, Relay board status leds, Zero or low sample flow – Teledyne 6200T - Sulfides Analyzer User Manual

Page 246

Advertising
background image

Troubleshooting

Model 6200T Total Sulfur Analyzer

246

Teledyne Analytical Instruments

12.2.3. Relay Board Status LEDs

The most important status LED on the relay board is the red I

2

C Bus watch-dog LED,

labeled D1 (or W/D), which indicates the health of the I

2

C communications bus. This

LED is located in the upper left-hand corner of the relay board when looking at the
electronic components. If D1 is blinking, then the other LEDs can be used in
conjunction with the DIAG menu I/O functions to test hardware functionality by
switching devices on and off and watching the corresponding LED turn on or off. The
LED only indicates that the logic signal for an output has been activated. If the output
driver (i.e. the relay or valve driver IC) is defective, then the LED will light up, but the
attached peripheral device will not turn on.

Table 12-3:

Relay Board Status LEDs

LED COLOR FUNCTION

FAULT

STATUS

INDICATED FAILURE(S)

D1 red

Watchdog Circuit; I

2

C

bus operation.

Continuously

ON or OFF

Failed/Halted CPU
Faulty Mother Board, Valve Driver board or
Relay PCA
Faulty Connectors/Wiring between Motherboard,
Valve Driver board or Relay PCA
Failed/Faulty +5 VDC Power Supply (PS1)

12.3. Gas Flow Problems

The standard analyzer has one main flow path. With the IZS option installed, there is a
second flow path through the IZS oven that runs whenever the IZS is on standby to
purge SO

2

from the oven chamber. The IZS flow is not measured so there is no reading

for it on the front panel display. The full flow diagrams of the standard configuration
(refer to Figure 3-18) and with options installed (refer to Figure 3-19 and Figure 3-20)
help in troubleshooting flow problems. In general, flow problems can be divided into
three categories:

 Flow is too high
 Flow is greater than zero, but is too low, and/or unstable
 Flow is zero (no flow)

When troubleshooting flow problems, it is essential to confirm the actual flow rate
without relying on the analyzer’s flow display. The use of an independent, external flow
meter to perform a flow check as described in Section 12.5.2 is essential.

12.3.1. Zero or Low Sample Flow

If the pump is operating but the unit reports a XXXX gas flow, do the following three
steps:

 Check for actual sample flow
 Check pressures
 Carry out a leak check

To check the actual sample flow, disconnect the sample tube from the sample inlet on
the rear panel of the instrument. Ensure that the unit is in basic SAMPLE mode. Place a
finger over the inlet and see if it gets sucked in by the vacuum or, more properly, use a
flow meter to measure the actual flow. If a proper flow of approximately 650 cm³/min

Advertising