6200T Total Sulfur Analyzer

Specifications

Teledyne Analytical Instruments

37

Table 2-4: M501TS Converter Specifications

Specification Value

Unit

Maximum Flow Rate

1000

cc/min

Nominal Flow Rate (CO

2

) 625

cc/min

Nominal Flow Rate (Air/N

2

) 450

cc/min

Maximum TS Concentration for specified
conversion efficiency

20 ppmv

Minimum Conversion Efficiency
(In CO

2

matrix)

H

2

S

COS, CS

2



98
90

%

Least Discernible Level (LDL)

See 6400T Manual

Operating Converter Temperature

1000

˚ C

Maximum Converter Temperature

1050

˚ C

Power

100-120/220-240 VAC
50/60 Hz, 440 watts

Weight

24 (11)

lbs. (kg)

Dimensions

7 x 17 x 22 (178 x 432 x 559)

inches (mm)

2.2. EPA Equivalency Designation

The 6200T Analyzer is designated as Reference Method Number EQSA-0495-100 as
per 40 CFR Part 53 when operated under the following conditions:

 Range: Any range from 50 parts per billion (ppb) to 10 parts per million

(ppm)

 Ambient temperature range of 5°C to 40°C
 Line voltage range of 100-120 VAC or 220-240 VAC, at 50 or 60 Hz
 Sample filter: Equipped with PTFE filter element in the internal filter

assembly

 Sample flow of 650 +/- 65 cm3/min
 Vacuum pump (internal) capable of 14" Hg Absolute pressure @ 1 slpm or

better

 Software settings:

Linearity

<± 0.1% CO

2

Temperature Coefficient

<± 0.01% CO

2

/°C

Rise and Fall Time

<60 seconds to 95%

1

As defined by the USEPA

2

Defined as twice the zero noise level by the USEPA

Advertising