Teledyne 6200T - Sulfides Analyzer User Manual

Page 70

Advertising
background image

Getting Started

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

70

ZERO CAL

Sample/Cal

Open to zero/span valve

Zero/Span

Open to ZERO AIR inlet

SPAN CAL

Sample/Cal

Open to zero/span valve

Zero/Span

Open to SPAN GAS inlet

The state of the IZS valves can also be controlled by any of the following means:

 Manually from the analyzer’s front panel by using the SIGNAL I/O

controls under the DIAG Menu (refer to Section 5.9.1),

 By activating the instrument’s AutoCal feature (refer to Section 9.8),
 Remotely by using the external digital control inputs (refer to Section 8.1.2

and Section 9.7.1),

 Remotely through the RS-232/485 serial I/O ports (refer to Appendix A-6

for the applicable commands), or

 Remotely via Ethernet

Note

The permeation tube is not included in the IZS Option and must be
ordered separately. Refer to Section 1.4 for permeation tube options.

3.3.2.5. Permeation Tube Heater

In order to keep the permeation rate constant, the IZS enclosure is heated to a constant
50 C (10° above the maximum operating temperature of the instrument). The IZS heater
is controlled by a precise PID (Proportional/Integral/Derivative) temperature control
loop. A thermistor measures the actual temperature and reports it to the CPU for control
feedback.
The IZS option includes an external zero air scrubber assembly that removes all SO

2

the

zero air source. The scrubber is filled with activated charcoal.

3.3.2.6. Span Gas Concentration Variation

Span gas is created when zero air passes over a permeation tube containing liquid SO

2

under high pressure, which slowly permeates through a PTFE membrane into the
surrounding air. The speed at which the SO

2

permeates the membrane is called the

effusion rate. The concentration of the span gas is determined by three factors:

 Size of the membrane: The larger the area of the membrane, the more

permeation occurs.

 Temperature of the SO

2

: Increasing the temperature of the increases the

pressure inside the tube and therefore increases the effusion rate.

 Flow rate of the zero air: If the previous two variables are constant, the

permeation rate of air into the zero air stream will be constant. Therefore, a
lower flow rate of zero air produces higher concentrations of SO

2

. The

6200T usually has a constant flow rate and a constant permeation rate;
hence, variations in concentration can be achieved by changing the IZS
temperature.

Advertising