6200T Total Sulfur Analyzer

Getting Started

Teledyne Analytical Instruments

55

CONTROL IN

A B C D E F U

+

SP

A

N

C

A

L

ZE

R

O

C

A

L

CONTROL IN

A B C D E F U

+

SP

A

N

C

A

L

ZE

R

O

C

A

L

-

+

5 VDC Power

Supply

Local Power Connections

External Power Connections

Figure 3-10:

Control Input Connector

Table 3-8: Control Input Signals

Input #

Status Definition

ON Condition

A

REMOTE ZERO CAL

The analyzer is placed in Zero Calibration mode. The mode field of the
display will read ZERO CAL R.

B

REMOTE SPAN CAL

The analyzer is placed in span calibration mode as part of performing a low
span (midpoint) calibration. The mode field of the display will read LO CAL
R

.

C, D, E & F SPARE

Digital Ground

The ground level from the analyzer’s internal DC power supplies (same as
chassis ground)

U

External Power input

Input pin for +5 VDC is required to activate pins A – F.

+

5 VDC output

Internally generated 5V DC power. To activate inputs A – F, place a jumper
between this pin and the “U” pin. The maximum amperage through this port
is 300 mA (combined with the analog output supply, if used).