6200t total sulfur analyzer setup menu, Analog output signal type and range span selection – Teledyne 6200T - Sulfides Analyzer User Manual

Page 121

Advertising
background image

6200T Total Sulfur Analyzer

Setup Menu

Teledyne Analytical Instruments

121

5.9.3.1. Analog Output Signal Type and Range Span Selection

To select an output signal type (DC Voltage or current) and level for one output channel,
activate the ANALOG I/O CONFIGURATION MENU from the DIAG Menu (refer
to Figure 5-16), then press:

DIAG AIO

CONC_OUT_1 RANGE: 5V

SET>

EDIT

EXIT

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

DIAG AIO

AOUTS CALIBRATED: NO

< SET

SET>

CAL

EXIT

DIAG AIO

CONC_OUT_1 RANGE: 5V

0.1V 1V 5V

10V

CURR

ENTR

EXIT

These buttons set

the signal level

and type for the

selected channel

DIAG AIO

CONC_OUT_1:5V,OVR,CAL

< SET SET>

EDIT

EXIT

Press SET> to select the
analog output channel to be
configured. Press EDIT to
continue

Pressing ENTR records the new setting

and returns to the previous menu.

Pressing EXIT ignores the new setting and

returns to the previous menu.

DIAG AIO

CONC_OUT_1 RANGE: 10V

0.1V 1V 5V

10V

CURR

ENTR EXIT

Figure 5-19:

DIAG – Analog I/O Configuration Menu