6200t total sulfur analyzer setup menu, Analog output signal type and range span selection – Teledyne 6200T - Sulfides Analyzer User Manual
Page 121
6200T Total Sulfur Analyzer
Setup Menu
Teledyne Analytical Instruments
121
5.9.3.1. Analog Output Signal Type and Range Span Selection
To select an output signal type (DC Voltage or current) and level for one output channel,
activate the ANALOG I/O CONFIGURATION MENU from the DIAG Menu (refer
to Figure 5-16), then press:
DIAG AIO
CONC_OUT_1 RANGE: 5V
SET>
EDIT
EXIT
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED: NO
< SET
SET>
CAL
EXIT
DIAG AIO
CONC_OUT_1 RANGE: 5V
0.1V 1V 5V
10V
CURR
ENTR
EXIT
These buttons set
the signal level
and type for the
selected channel
DIAG AIO
CONC_OUT_1:5V,OVR,CAL
< SET SET>
EDIT
EXIT
Press SET> to select the
analog output channel to be
configured. Press EDIT to
continue
Pressing ENTR records the new setting
and returns to the previous menu.
Pressing EXIT ignores the new setting and
returns to the previous menu.
DIAG AIO
CONC_OUT_1 RANGE: 10V
0.1V 1V 5V
10V
CURR
ENTR EXIT
Figure 5-19:
DIAG – Analog I/O Configuration Menu