6200T Total Sulfur Analyzer
Instrument Test and Calibration Record
Teledyen Analytical instruments
343
Table 17-2.
Test and Calibration Values for 6200TU
TEST
PARAMETERS
OBSERVED
VALUE
UNITS ACCEPTABLE
VALUE(S)
RANGE
PPB
5 - 20,000
STAB1
PPB
0.05 ppb with zero air
STAB2
PPB
0.1 ppb with zero air
PRESS
“
HG
ambient ± 2
SAMPLE FL
CC / MIN
650 cm3/min ± 10%
PMT
mV
-20 TO 150 mV with zero air
UV LAMP
mV
2000 - 4800
STR LGT
PPB
< 25
DRK PMT
MV
200 - 325
DRK LMP
MV
-50 - 200
SLOPE
1.0 ± 0.3
OFFSET
MV
<
250
HVPS
V
≈ 400 to 900
RCELL TEMP
o
C
50 ± 1º
BOX TEMP
o
C ambient
+ ~ 5
PMT TEMP
o
C
7 ± 2
IZS TEMP (option)
o
C
50 ± 1
Electric Test
PMT Volts
MV
1000 +/-200
TS Conc
PPB
500 +/- 100
Optic Test
PMT Volts
MV
1000 +/- 200
TS Conc
PPB
500 +/- 100