Teledyne 6200T - Sulfides Analyzer User Manual

Page 343

Advertising
background image

6200T Total Sulfur Analyzer

Instrument Test and Calibration Record

Teledyen Analytical instruments

343

Table 17-2.

Test and Calibration Values for 6200TU

TEST

PARAMETERS

OBSERVED

VALUE

UNITS ACCEPTABLE

VALUE(S)

RANGE

PPB

5 - 20,000

STAB1

PPB

0.05 ppb with zero air

STAB2

PPB

0.1 ppb with zero air

PRESS

HG

ambient ± 2

SAMPLE FL

CC / MIN

650 cm3/min ± 10%

PMT

mV

-20 TO 150 mV with zero air

UV LAMP

mV

2000 - 4800

STR LGT

PPB

< 25

DRK PMT

MV

200 - 325

DRK LMP

MV

-50 - 200

SLOPE

1.0 ± 0.3

OFFSET

MV

<

250

HVPS

V

≈ 400 to 900

RCELL TEMP

o

C

50 ± 1º

BOX TEMP

o

C ambient

+ ~ 5

PMT TEMP

o

C

7 ± 2

IZS TEMP (option)

o

C

50 ± 1

Electric Test

PMT Volts

MV

1000 +/-200

TS Conc

PPB

500 +/- 100

Optic Test

PMT Volts

MV

1000 +/- 200

TS Conc

PPB

500 +/- 100