6200T Total Sulfur Analyzer

Troubleshooting

Teledyen Analytical Instruments

281

Figure 12-21: Quartz Tube Cavity for Thermocouple

11. The thermocouple should slide into the Heater Block and into the indentation of

the quartz.

12. Align the thermocouple with this cavity and carefully push the thermocouple all

the way into the cavity until it comes to a stop, which is the end of the cavity of
the quartz tube.

13. The thermocouple should now be properly seated in the cavity of the quartz

tube. Refer to Figure

Figure 12-22: Thermocouple Installed

14. Reconnect the Teflon fitting that was removed earlier from the end of the quartz

tube. Take care not to put any stress on the quartz tube as the Teflon fitting is
tightened.

15. Clean the chassis where the Tie-Wrap Hold-Down will be placed (alcohol is

recommended), and place the Tie-Wrap Hold-Down as shown in the Figure .