Pmt optical filter – Teledyne 6200T - Sulfides Analyzer User Manual

Page 292

Advertising
background image

Principles of Operation

Model 6200T Total Sulfur Analyzer

292

Teledyne Analytical Instruments

13.1.7.2. PMT Optical Filter

The PMT used in the 6200T reacts to a wide spectrum of light which includes much of
the visible spectrum and most of the UV spectrum. Even though the 214 nm light used
to excite the SO

2

is focused away from the PMT, some of it scatters in the direction of

the PMT as it interacts with the sample gas. A second optical band pass filter placed
between the sample chamber (refer to
Figure 13-2) and the PMT strips away light outside of the fluorescence spectrum of
decaying SO

2

* (refer to

Figure 13-5) including reflected UV form the source lamp and other stray light.

LAM

P

OUT

P

UT

(A

rbit

rary Un

tis)

10

5

10

4

10

2

10

3

10

1

1

100

200

300

400 500

WAVELENGTH (nm)

0

21

3.9

330

.3

SO

2

* FLUORESCENT

SPECTRUM

PMT OPTICAL FILTER

BANDWIDTH

Figure 13-5: PMT Optical Filter Bandwidth