Chassis – Teledyne 6200T - Sulfides Analyzer User Manual
Page 68
Getting Started
Model 6200T Total Sulfur Analyzer
Teledyne Analytical Instruments
68
3.3.2.3. Pneumatic Layout for Zero/Span Valves Option
Figure 3-19 shows the internal, pneumatic connections for a 6200T with the zero/span
valve option installed.
FLOW
SENSOR
FLOW / PRESSURE
SENSOR PCA
SAMPLE
PRESSURE
SENSOR
V
ACUUM
M
ANI
F
O
L
D
FLOW
CONTROL
ASSY
EXHAUST TO OUTER
LAYER OF KICKER
SAMPLE/CAL
VALVE
SAMPLE GAS
INLET
ZERO AIR INLET
SPAN 1 INLET
ZERO/SPAN
VALVE
SAMPLE FILTER
Chassis
EXHAUST GAS
OUTLET
KICKER EXHAUST
TO PUMP
HYDROCARBON
SCRUBBER
(KICKER)
UV
LAMP
PMT
SAMPLE
CHAMBER
PUMP
Figure 3-19:
Pneumatic Layout with Zero/Span Valves Option
Table 3-10 describes the state of each valve during the analyzer’s various operational
modes.
Table 3-10: Zero/Span and Sample/Cal Valve Operating States
MODE VALVE
CONDITION
SAMPLE
Sample/Cal
Open to SAMPLE inlet
Zero/Span
Open to ZERO AIR inlet
ZERO CAL
Sample/Cal
Open to zero/span inlet
Zero/Span
Open to ZERO AIR inlet
SPAN CAL
Sample/Cal
Open to zero/span inlet
Zero/Span
Open to SPAN GAS inlet
The state of the zero/span valves can also be controlled by any of the following means:
manually from the analyzer’s front panel by using the SIGNAL I/O
controls located within the DIAG Menu (refer to Section 5.9.1)
by activating the instrument’s AutoCal feature (refer to Section 9.8)
remotely by using the external digital control inputs (refer to Section 8.1.2
and Section 9.7.1)