Chassis – Teledyne 6200T - Sulfides Analyzer User Manual

Page 68

Advertising
background image

Getting Started

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

68

3.3.2.3. Pneumatic Layout for Zero/Span Valves Option

Figure 3-19 shows the internal, pneumatic connections for a 6200T with the zero/span
valve option installed.

FLOW

SENSOR

FLOW / PRESSURE

SENSOR PCA

SAMPLE

PRESSURE

SENSOR

V

ACUUM

M

ANI

F

O

L

D

FLOW
CONTROL
ASSY

EXHAUST TO OUTER

LAYER OF KICKER

SAMPLE/CAL

VALVE

SAMPLE GAS

INLET

ZERO AIR INLET

SPAN 1 INLET

ZERO/SPAN
VALVE

SAMPLE FILTER

Chassis

EXHAUST GAS

OUTLET

KICKER EXHAUST

TO PUMP

HYDROCARBON

SCRUBBER

(KICKER)

UV
LAMP

PMT

SAMPLE

CHAMBER

PUMP

Figure 3-19:

Pneumatic Layout with Zero/Span Valves Option

Table 3-10 describes the state of each valve during the analyzer’s various operational
modes.

Table 3-10: Zero/Span and Sample/Cal Valve Operating States

MODE VALVE

CONDITION

SAMPLE

Sample/Cal

Open to SAMPLE inlet

Zero/Span

Open to ZERO AIR inlet

ZERO CAL

Sample/Cal

Open to zero/span inlet

Zero/Span

Open to ZERO AIR inlet

SPAN CAL

Sample/Cal

Open to zero/span inlet

Zero/Span

Open to SPAN GAS inlet

The state of the zero/span valves can also be controlled by any of the following means:

 manually from the analyzer’s front panel by using the SIGNAL I/O

controls located within the DIAG Menu (refer to Section 5.9.1)

 by activating the instrument’s AutoCal feature (refer to Section 9.8)
 remotely by using the external digital control inputs (refer to Section 8.1.2

and Section 9.7.1)