Setup – comm: communications ports, Id (instrument identification), 6200t total sulfur analyzer setup menu – Teledyne 6200T - Sulfides Analyzer User Manual

Page 111

Advertising
background image

6200T Total Sulfur Analyzer

Setup Menu

Teledyne Analytical Instruments

111

5.7. SETUP – COMM: Communications Ports

This section introduces the communications setup menu; Section 6 provides the setup
instructions and operation information. Press SETUP>ENTR>MORE>COMM to arrive
at the communications menu.

SAMPLE RANGE = 500.000 PPB SO2 =XXX.X

< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS DIAG

EXIT

SETUP X.X

COMMUNICATIONS MENU

ID INET

COM1 COM2

EXIT

If the default password 818
is replaced by 000, then
Password Protection has
been enabled. Refer to
SETUP: PASS

.

Figure 5-13:

SETUP – COMM Menu

5.7.1. ID (Instrument Identification)

Press ID to display and/or change the Machine ID, which must be changed to a unique
identifier (number) when more than one instrument of the same model is used in a
multidrop configuration (0) or when applying MODBUS protocol (Section 6.6.1). The
default ID is the same as the model number; for the Model 6200T, the ID is 0100. Press
any button(s) in the MACHINE ID menu (Figure 5-14) until the Machine ID Parameter
field displays the desired identifier.