6200T Total Sulfur Analyzer
Setup Menu
Teledyne Analytical Instruments
111
5.7. SETUP – COMM: Communications Ports
This section introduces the communications setup menu; Section 6 provides the setup
instructions and operation information. Press SETUP>ENTR>MORE>COMM to arrive
at the communications menu.
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG
EXIT
SETUP X.X
COMMUNICATIONS MENU
ID INET
COM1 COM2
EXIT
If the default password 818
is replaced by 000, then
Password Protection has
been enabled. Refer to
SETUP: PASS
.
Figure 5-13:
SETUP – COMM Menu
5.7.1. ID (Instrument Identification)
Press ID to display and/or change the Machine ID, which must be changed to a unique
identifier (number) when more than one instrument of the same model is used in a
multidrop configuration (0) or when applying MODBUS protocol (Section 6.6.1). The
default ID is the same as the model number; for the Model 6200T, the ID is 0100. Press
any button(s) in the MACHINE ID menu (Figure 5-14) until the Machine ID Parameter
field displays the desired identifier.