Calibration with remote contact closures – Teledyne 6200T - Sulfides Analyzer User Manual

Page 200

Advertising
background image

Calibration Procedures

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

200

9.7. Manual Calibration in DUAL or AUTO Reporting Range

Modes

When the analyzer is in either Dual or Auto Range modes the user must run a separate
calibration procedure for each range. After pressing the CAL, CALZ or CALS buttons
the user is prompted for the range that is to be calibrated as seen in the CALZ example
below:

Continue Calibration as per

Standard Procedure

SAMPLE RANGE = 500.000 PPB SO2 =XXX.X

< TST TST > CAL

CALZ

CALS

SETUP

ZERO CAL M

RANGE = 500.000 PPB SO2 =XXX.X

< TST TST >

ZERO

CONC

EXIT

WAIT 10

MINUTES

Or until the

reading

stabilizes and

the ZERO button

is displayed

SAMPLE

RANGE TO CAL: LOW

LOW

HIGH

ENTR SETUP

SAMPLE

RANGE TO CAL: HIGH

LOW HIGH

ENTR

SETUP

Analyzer enters

ZERO CAL

Mode

Refer to Table 6-1 for

Z/S Valve States

during this operating

mode.

Figure 9-11:

Manual Calibration in Dual/Auto Reporting Range Modes

Once this selection is made, the calibration procedure continues as previously described
in Sections 7.2 through 7.6. The other range may be calibrated by starting over from the
main SAMPLE display.

9.7.1. Calibration With Remote Contact Closures

Contact closures for controlling calibration and calibration checks are located on the rear
panel CONTROL IN connector. Instructions for setup and use of these contacts can be
found in Section 8.1.2.
When the appropriate contacts are closed for at least 5 seconds, the instrument switches
into zero, low span or high span mode and the internal zero/span valves will be
automatically switched to the appropriate configuration. The remote calibration contact
closures may be activated in any order. It is recommended that contact closures remain
closed for at least 10 minutes to establish a reliable reading; the instrument will stay in
the selected mode for as long as the contacts remain closed.
If contact closures are used in conjunction with the analyzer’s AutoCal (refer to Section
9.8) feature and the AutoCal attribute CALIBRATE is enabled, the 6200T will not re-

Advertising