Calibration with remote contact closures – Teledyne 6200T - Sulfides Analyzer User Manual
Page 200
Calibration Procedures
Model 6200T Total Sulfur Analyzer
Teledyne Analytical Instruments
200
9.7. Manual Calibration in DUAL or AUTO Reporting Range
Modes
When the analyzer is in either Dual or Auto Range modes the user must run a separate
calibration procedure for each range. After pressing the CAL, CALZ or CALS buttons
the user is prompted for the range that is to be calibrated as seen in the CALZ example
below:
Continue Calibration as per
Standard Procedure
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
CALZ
CALS
SETUP
ZERO CAL M
RANGE = 500.000 PPB SO2 =XXX.X
< TST TST >
ZERO
CONC
EXIT
WAIT 10
MINUTES
Or until the
reading
stabilizes and
the ZERO button
is displayed
SAMPLE
RANGE TO CAL: LOW
LOW
HIGH
ENTR SETUP
SAMPLE
RANGE TO CAL: HIGH
LOW HIGH
ENTR
SETUP
Analyzer enters
ZERO CAL
Mode
Refer to Table 6-1 for
Z/S Valve States
during this operating
mode.
Figure 9-11:
Manual Calibration in Dual/Auto Reporting Range Modes
Once this selection is made, the calibration procedure continues as previously described
in Sections 7.2 through 7.6. The other range may be calibrated by starting over from the
main SAMPLE display.
9.7.1. Calibration With Remote Contact Closures
Contact closures for controlling calibration and calibration checks are located on the rear
panel CONTROL IN connector. Instructions for setup and use of these contacts can be
found in Section 8.1.2.
When the appropriate contacts are closed for at least 5 seconds, the instrument switches
into zero, low span or high span mode and the internal zero/span valves will be
automatically switched to the appropriate configuration. The remote calibration contact
closures may be activated in any order. It is recommended that contact closures remain
closed for at least 10 minutes to establish a reliable reading; the instrument will stay in
the selected mode for as long as the contacts remain closed.
If contact closures are used in conjunction with the analyzer’s AutoCal (refer to Section
9.8) feature and the AutoCal attribute CALIBRATE is enabled, the 6200T will not re-