Usb port – Teledyne 6200T - Sulfides Analyzer User Manual
Page 148
Communications Setup
Model 6200T Total Sulfur Analyzer
Teledyne Analytical Instruments
148
change this name (particularly if you have more than one 6200T analyzer on your
network), access the COMMUNICATIONS Menu (refer to Figure 5-13), then press:
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
COM2
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
HOSTNAME: T100
<SET
EDIT
EXIT
SETUP X.X
HOSTNAME: T100
<CH CH> INS DEL [?]
ENTR EXIT
SETUP X.X
HOSTNAME: T100-FIELD1
<SET
EDIT
EXIT
Continue pressing SET> UNTIL …
SETUP X.X
INITIALIZING INET 0%
…
INITIALIZING INET 100%
SETUP X.X
INITIALIZATI0N SUCCEEDED
SETUP X.X
INITIALIZATION FAILED
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
COM2
EXIT
Contact your IT Network
Administrator
Use these buttons (See
Table 5-3) to edit HOSTNAME
Figure 6-6: COMM – Change Hostname
Table 6-4:
Hostname Editing Button Functions
Button Function
<CH
Moves the cursor one character to the left.
CH>
Moves the cursor one character to the right.
INS
Inserts a character before the cursor location.
DEL
Deletes a character at the cursor location.
[?]
Press this button to cycle through the range of numerals and characters
available for insertion.:
0-9, A-Z, space ’ ~ !
# $ % ^ & * ( ) - _ = +[ ] { } < >\ | ; : , . / ?
ENTR
Accepts the new setting and returns to the previous menu.
EXIT
Ignores the new setting and returns to the previous menu.
Buttons only appear when applicable.
6.5.3. USB Port
Using the USB port disallows use of the rear panel COM2 port except when using the
COM2 port for multidrop communication. USB configuration requires matching the
baud rates of the instrument and the PC to which it is connected. To view or change the
instrument baud rate: