Checking for light leaks – Teledyne 6200T - Sulfides Analyzer User Manual

Page 231

Advertising
background image

6200T Total Sulfur Analyzer

Maintenance

Teledyen Analytical Instruments

231

Gas Line fitting

Spring

Sintered Filter

O-Ring

Critical Flow Orifice

O-Ring

Vacuum Manifold

Figure 11-2:

Critical Flow Orifice Assembly

5. Take out the components of the assembly: a spring, a sintered filter, two O-rings

and the critical flow orifice.

6.

You may need to use a scribe or pressure from the vacuum port to get the parts
out

of the manifold.

7. Discard the two O-rings and the sintered filter.

8. Replace the critical flow orifice.

9. Let the part dry.

10. Re-assemble the parts as shown in Figure 11-2 using a new filter and o-rings.

11. Reinstall the NPT fitting and connect all tubing.

12. Power up the analyzer and allow it to warm up for 60 minutes.

13. Perform a leak check (refer to Section 11.3.6).

11.3.5. Checking for Light Leaks

When re-assembled after maintenance, repair or improper operation, the 6200T can
develop small leaks around the PMT, allowing stray light from the analyzer
surroundings into the PMT housing. To find light leaks, follow the below procedures:

CAUTION

This procedure must be carried out by qualified personnel, as it must be
performed while the analyzer is powered up and running and its cover
removed.

Advertising