Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

78

Table 3-13 lists brief descriptions of the warning messages that may occur during start
up for 6200T analyzers with no options installed.

Table 3-13:

Possible Startup Warning Messages – 6200T Analyzers w/o Options

Message Meaning

ANALOG CAL WARNING

The instrument's A/D circuitry or one of its analog outputs is not calibrated.

BOX TEMP WARNING

The temperature inside the 6200T chassis is outside the specified limits.

CANNOT DYN SPAN

2

Remote span calibration failed while the dynamic span feature was set to
turned on.

CANNOT DYN ZERO

3

Remote zero calibration failed while the dynamic zero feature was set to turned
on.

CONFIG INITIALIZED

Configuration was reset to factory defaults or was erased.

DARK CAL WARNING

Dark offset above limit specified indicating that too much stray light is present in
the sample chamber.

DATA INITIALIZED

DAS data storage was erased.

HVPS WARNING

High voltage power supply for the PMT is outside of specified limits.

PMT DET WARNING

PMT detector output is outside of operational limits.

PMT TEMP WARNING

PMT temperature is outside of specified limits.

RCELL TEMP WARNING

Sample chamber temperature is outside of specified limits.

REAR BOARD NOT DET

CPU unable to communicate with motherboard.

RELAY BOARD WARN

CPU is unable to communicate with the relay PCA.

SAMPLE FLOW WARN

The flow rate of the sample gas is outside the specified limits.

SAMPLE PRESS WARN

Sample gas pressure outside of operational parameters.

SYSTEM RESET

1

The computer was rebooted.

UV LAMP WARNING

The UV lamp intensity measured by the reference detector reading too low or
too high.

1

Clears 45 minutes after power up.

2

Clears the next time successful zero calibration is performed.

3

Clears the next time successful span calibration is performed.

MTS501 Converter Messages

UUUU

The thermocouple in the converter has failed.

Table 3-14 lists brief descriptions of the warning messages that may occur during start