Troubleshooting

Model 6200T Total Sulfur Analyzer

282

Teledyne Analytical Instruments

Figure 12-23: Tie-Wrap Hold-Down Location

16. Form the Thermocouple wire so that it rests in the cavity with little movement.

17. Connect the (2) wires of the thermocouple to the Temperature Controller. The

Black wire should be connected to Pin 1 and the Red wire should be connected
to Pin 2. (If the wires are of any other color, STOP. Get the correct part from
TAI Sales or call Customer Service).

18. At this point, all connections have been made, both electrically and

pneumatically. A leak check should be performed on the Converter to verify
that all connections are leak free. If a leak is detected, the leak should be
resolved before continuing.

19. Install the inner cover of the Heater Block and secure with the (3) nuts. Close

the Front Panel and secure with the (4) nuts. Install the top cover on the
Converter chassis.

20. The Converter is now ready for the application of power. You will be looking

for an indication from the temperature controller that it is functioning correctly
and driving the heater to the desired “set” temperature. Apply power now.

21. Check the Temp Controller to be sure that it knows which type of

Thermocouple it has in it. Follow the directions in Section 3.4.1.3 of this
document, paying particular attention to the settings in Table 2-5 to be sure that
the temp controller is set properly.