Figure 5-15: setup – vars menu – Teledyne 6200T - Sulfides Analyzer User Manual

Page 114

Advertising
background image

Setup Menu

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

114

To access and navigate the VARS menu, use the following button sequence.

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SAMPLE* RANGE = 500.000 PPB SO2 =X.XXX

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS

DIAG

EXIT

SETUP X.X

0 ) DAS_HOLD_OFF=

15.0 Minutes

NEXT

JUMP

EDIT

PRNT EXIT

SETUP X.X

1 ) TPC_ENABLE=ON

PREV

NEXT

JUMP

EDIT

PRNT EXIT

SETUP X.X

ENTER VARS PASS: 818

8 1 8

ENTR

EXIT

SETUP X.X

DAS_HOLD_OFF=15.0 Minutes

1 5 .0

ENTR EXIT

SETUP X.X

2)RCELL_SET=

50.0 DegC

PREV

NEXT

JUMP

EDIT

PRNT EXIT

SETUP X.X

3) IZS_SET=

50.0 DegC

PREV

NEXT

JUMP

EDIT

PRNT EXIT

DO NOT

change

theses set-points

unless

specifically

instructed to by

TAI Customer

Service.

SETUP X.X

CLOCK_ADJ=0 Sec/Day

+ 0 0

ENTR EXIT

SETUP X.X

7) CLOCK_ADJ=0 Sec/Day

PREV

NEXT

JUMP

EDIT

PRNT EXIT

EXIT

ignores the new setting.

ENTR

accepts the new setting.

SETUP X.X

TPC_ENABLE=ON


ON

ENTR EXIT

SETUP X.X

4 ) DYN_ZERO=ON

PREV

NEXT

JUMP

EDIT

PRNT EXIT

SETUP X.X

DYN_ZERO=ON


ON

ENTR EXIT

Toggle to change setting

Toggle to change setting

SETUP X.X

5) DYN_SPAN=ON

PREV

NEXT

JUMP

EDIT

PRNT EXIT

SETUP X.X

DYN_SPAN=ON


ON

ENTR EXIT

Toggle to change setting

Toggle to change setting

Toggle to change setting

SETUP X.X

CONC_PRECUISION : 3

AUTO 0 1 2 3 4 ENTR EXIT

SETUP X.X

6) CONC_PRECISION : 3

PREV

NEXT

JUMP

EDIT

PRNT EXIT

Toggle to change setting

Figure 5-15:

SETUP – VARS Menu

Advertising