Editing das data channels, 6200t total sulfur analyzer das and aicom – Teledyne 6200T - Sulfides Analyzer User Manual

Page 163

Advertising
background image

6200T Total Sulfur Analyzer

DAS and AICOM

Teledyne Analytical Instruments

163

7.2.2. Editing DAS Data Channels

Although DAS configuration is most conveniently done through the AICOM remote
control program (refer to Section 6.12.2.8), the following illustrations shows how to edit
DAS channels using the analyzer’s front panel control buttons.

Main Data Acquisition Menu

Edit Data Channel Menu

SAMPLE RANGE = 500.000 PPB SO2 =XXX.X

< TST TST > CAL

SETUP

SETUP X.X

0) CONC1: ATIMER, 2, 4032, RS-232


PREV NEXT

INS DEL EDIT PRNT EXIT

SETUP X.X

PRIMARY SETUP MENU


CFG

DAS

RNGE

PASS CLK MORE

EXIT

SAMPLE

ENTER SETUP PASS : 818


8 1 8

ENTR

EXIT

EXIT

will return to the

previous SAMPLE

display.

SETUP X.X

DATA ACQUISITION

VIEW

EDIT

EXIT

Moves the

display up &

down the list of
Data Channels

Inserts

a new Data

Channel into the list

BEFORE the Channel

currently being displayed

Deletes

the Data

Channel currently

being displayed

Exports the

configuration of all

data channels to

RS-232 interface.

Exits

to the Main

Data Acquisition

Menu

SETUP X.X

NAME:CONC1

<SET SET>

EDIT PRNT

EXIT

Moves the display

between the

PROPERTIES for this

data channel.

Reports the configuration of current

data channels to the RS-232 ports.

Exits

returns to the

previous Menu

Allows to edit the channel name, see next menu button.

Figure 7-3: DAS – Editing DAS Data Channels