Setup Menu

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

134

5.9.6. Lamp Calibration

An important factor in accurately determining SO

2

concentration is the amount of UV

light available to transform the SO

2

into SO

2

* (refer to Section 13.1.1). The 6200T

compensates for variations in the intensity of the available UV light by adjusting the SO

2

concentration calculation using a ratio (LAMP RATIO)that results from dividing the
current UV lamp (UV LAMP) intensity by a value stored in the CPU’s memory
(LAMP_CAL). Both LAMP Ration and UV Lamp are test functions viewable from the
instruments front panel.
To cause the analyzer to measure and record a value for LAMP_CAL, access the Signal
I/O from the DIAG Menu (refer to Figure 5-16), then press:

DIAG

SIGNAL I / O

NEXT

ENTR

EXIT

Repeat Pressing NEXT until . .

.

DIAG

LAMP CALIBRATION

PREV NEXT

ENTR

EXIT

DIAG FCAL

LAMP CAL VALUE:4262.4 mV


4 2 6 2 .4 ENTR EXIT

ENTR accepts the

new value

EXIT

ignores the new

value

The value displayed is the

current output of the UV

source reference detector

Figure 5-32:

DIAG – Lamp Calibration