6200T Total Sulfur Analyzer

Setup Menu

Teledyne Analytical Instruments

125

5.9.3.3. Manual Analog Output Calibration and Voltage Adjustment

For highest accuracy, the voltages of the analog outputs can be manually calibrated.
Calibration is done through the instrument software with a voltmeter connected across
the output terminals (refer to Figure 5-23). Adjustments are made using the control
buttons by setting the zero-point first and then the span-point (refer to Table 5-7).
The software allows this adjustment to be made in 100, 10 or 1 count increments.

Table 5-7: Voltage Tolerances for Analog Output Calibration

FULL SCALE

ZERO TOLERANCE

SPAN VOLTAGE

SPAN TOLERANCE

0.1 VDC

±0.0005V

90 mV

±0.001V

1 VDC

±0.001V

900 mV

±0.001V

5 VDC

±0.002V

4500 mV

±0.003V

10 VDC

±0.004V

4500 mV

±0.006V

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

Outputs configured for 0.1V full scale should always be calibrated
manually.

V

+DC Gnd

V OUT +

V OUT -

V IN +

V IN -

Recording

Device

ANALYZER

See Table 3-1 for

pin assignments

of Analog Out

connector on the

rear panel

Figure 5-23:

Setup for Calibrating Analog Outputs