5.9.3.3. Manual Analog Output Calibration and Voltage Adjustment
For highest accuracy, the voltages of the analog outputs can be manually calibrated.
Calibration is done through the instrument software with a voltmeter connected across
the output terminals (refer to Figure 5-23). Adjustments are made using the control
buttons by setting the zero-point first and then the span-point (refer to Table 5-7).
The software allows this adjustment to be made in 100, 10 or 1 count increments.
Table 5-7: Voltage Tolerances for Analog Output Calibration
FULL SCALE
ZERO TOLERANCE
SPAN VOLTAGE
SPAN TOLERANCE
0.1 VDC
±0.0005V
90 mV
±0.001V
1 VDC
±0.001V
900 mV
±0.001V
5 VDC
±0.002V
4500 mV
±0.003V
10 VDC
±0.004V
4500 mV
±0.006V
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
Outputs configured for 0.1V full scale should always be calibrated
manually.
V
+DC Gnd
V OUT +
V OUT -
V IN +
V IN -
Recording
Device
ANALYZER
See Table 3-1 for
pin assignments
of Analog Out
connector on the
rear panel
Figure 5-23:
Setup for Calibrating Analog Outputs