Rs-232 report function, Compact report, Starting date – Teledyne 6200T - Sulfides Analyzer User Manual
Page 171: Disabling/enabling data channels, 6200t total sulfur analyzer das and aicom
6200T Total Sulfur Analyzer
DAS and AICOM
Teledyne Analytical Instruments
171
7.2.7. RS-232 Report Function
The M DAS can automatically report data to the communications ports, where they can
be captured with a terminal emulation program or simply viewed by the user.
To enable automatic COMM port reporting, follow the instruction shown in Figure 7-3,
then press:
Edit Data Channel Menu
SETUP X.X
0) CONC: ATIMER, 2, 4032, R
PREV NEXT INS DEL
EDIT
PRNT
EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
EXIT
SETUP X.X
RS-232 REPORT: OFF
<SET SET>
EDIT
EXIT
ENTR
accepts the new
setting and returns to the
previous menu.
EXIT
ignores the new setting
and returns to the previous
menu.
SETUP X.X
RS-232 REPORT: OFF
OFF
ENTR EXIT
Toggle to turn
reporting ON or OFF
Press SET> until…
Figure 7-10:
DAS – RS-232 Report Function
7.2.8. Compact Report
When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead
of reporting each parameter in one channel on a separate line, up to five parameters are
reported in one line, instead.
7.2.9. Starting Date
This option allows a user to specify a starting date for any given channel in case the user
wants to start data acquisition only after a certain time and date. If the Starting Date is
in the past, the DAS ignores this setting.
7.2.10. Disabling/Enabling Data Channels
Data channels can be temporarily disabled, which can reduce the read/write wear on the
disk-on-module. The ALL_01 channel of the 6200T, for example, is disabled by default.