Rs-232 report function, Compact report, Starting date – Teledyne 6200T - Sulfides Analyzer User Manual

Page 171: Disabling/enabling data channels, 6200t total sulfur analyzer das and aicom

Advertising
background image

6200T Total Sulfur Analyzer

DAS and AICOM

Teledyne Analytical Instruments

171

7.2.7. RS-232 Report Function

The M DAS can automatically report data to the communications ports, where they can
be captured with a terminal emulation program or simply viewed by the user.
To enable automatic COMM port reporting, follow the instruction shown in Figure 7-3,
then press:


Edit Data Channel Menu

SETUP X.X

0) CONC: ATIMER, 2, 4032, R


PREV NEXT INS DEL

EDIT

PRNT

EXIT

Exits to the main

Data Acquisition

menu

SETUP X.X

NAME:CONC


<SET

SET>

EDIT

PRINT

EXIT

SETUP X.X

RS-232 REPORT: OFF


<SET SET>

EDIT

PRINT

EXIT

ENTR

accepts the new

setting and returns to the

previous menu.

EXIT

ignores the new setting

and returns to the previous

menu.

SETUP X.X

RS-232 REPORT: OFF

OFF

ENTR EXIT

Toggle to turn

reporting ON or OFF

Press SET> until…

Figure 7-10:

DAS – RS-232 Report Function

7.2.8. Compact Report

When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead
of reporting each parameter in one channel on a separate line, up to five parameters are
reported in one line, instead.

7.2.9. Starting Date

This option allows a user to specify a starting date for any given channel in case the user
wants to start data acquisition only after a certain time and date. If the Starting Date is
in the past, the DAS ignores this setting.

7.2.10. Disabling/Enabling Data Channels

Data channels can be temporarily disabled, which can reduce the read/write wear on the
disk-on-module. The ALL_01 channel of the 6200T, for example, is disabled by default.