6200T Total Sulfur Analyzer

Setup Menu

Teledyne Analytical Instruments

135

5.9.7. Pressure Calibration

A sensor at the exit of the sample chamber continuously measures the pressure of the
sample gas. This data is used to compensate the final SO

2

concentration calculation for

changes in atmospheric pressure when the instrument’s TPC feature is turned on (refer
to Section 10.7.3) and is stored in the CPU’s memory as the test function PRES (also
viewable via the front panel).
Ensure to use a barometer that measures actual barometric pressure.
To cause the analyzer to measure and record a value for PRES, access the Signal I/O
from the DIAG Menu (refer to Figure 5-16), then press:



DIAG

SIGNAL I / O

NEXT

ENTR

EXIT

Repeat Pressing NEXT until . .

.

DIAG

PRESSURE CALIBRATION

PREV NEXT

ENTR

EXIT

DIAG PCAL

ACTUAL PRES :27.20 IN-HG-A


2 7 .2 0 ENTR EXIT

ENTR accepts the

new value

EXIT

ignores the new

value

Adjust these values until the

displayed pressure equals the

pressure measured by the

independent pressure meter.

Figure 5-33:

DIAG – Pressure Calibration