Number of records, 6200t total sulfur analyzer das and aicom, Teledyne analytical instruments 169 – Teledyne 6200T - Sulfides Analyzer User Manual
Page 169: Figure 7-8: das – define the report period
6200T Total Sulfur Analyzer
DAS and AICOM
Teledyne Analytical Instruments
169
To define the REPORT PERIOD, follow the instruction shown in Figure 7-3, then
press:
Edit Data Channel Menu
SETUP X.X
0) CONC: ATIMER, 2, 4032, R
PREV NEXT INS DEL
EDIT
PRNT
EXIT
Exits to the main
Data Acquisition
menu.
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
EXIT
Press SET> until you reach REPORT PERIOD …
SETUP X.X
REPORT PERIODD:DAYS:0
0 0 0
ENTR EXIT
ENTR
accepts the new string and
returns to the previous menu.
EXIT
ignores the new string and
returns to the previous menu.
Press buttons to set amount of
time between reports, in hours
(HH) and/or minutes (MM)
(max: 23:59). 01:00 sets a
report to be made every hour.
Set the number of days
between reports (0-365).
IIf at any time an invalidl entry is selected (e.g., days > 366)
the ENTR button will disappear from the display.
SETUP X.X
REPORT PERIODD:TIME:01:00
0 1 0 0
ENTR EXIT
SETUP X.X
REPORT PERIOD:000:00:05
<SET SET>
EDIT
EXIT
Use PREV and NEXT to
scroll to the data channel
to be edited.
Figure 7-8: DAS – Define the Report Period
7.2.6. Number of Records
The number of data records in the 6200T is limited to about a cumulative one million
data points in all channels (one megabyte of space on the disk-on-module). However,
the actual number of records is also limited by the total number of parameters and
channels and other settings in the DAS configuration. Every additional data channel,
parameter, number of samples setting etc. will reduce the maximum amount of data
points somewhat. In general, however, the maximum data capacity is divided amongst
all channels (max: 20) and parameters (max: 50 per channel).
The DAS will check the amount of available data space and prevent the user from
specifying too many records at any given point. If, for example, the DAS memory space
can accommodate 375 more data records, the ENTR button will disappear when trying
to specify more than that number of records. This check for memory space may also
make an upload of a DAS configuration with AICOM or a Terminal program fail, if the
combined number of records would be exceeded. In this case, it is suggested to either try
from the front panel what the maximum number of records can be or use trial-and-error
in designing the DAS script or calculate the number of records using the DAS or
AICOM manuals. To set the number of records for one channel from the front panel,
press SETUP-DAS-EDIT-ENTR and the following control button sequence.