Sensor module – Teledyne 6200T - Sulfides Analyzer User Manual

Page 306

Advertising
background image

Principles of Operation

Model 6200T Total Sulfur Analyzer

306

Teledyne Analytical Instruments

In the unlikely event that the flash chip should fail, the analyzer will continue to operate
with just the DOM. However, all configuration information will be lost, requiring that
the unit be recalibrated.

13.5.2. Sensor Module

Electronically, the 6200T sensor module is a group of components that: create the UV
light that initiates the fluorescence reaction between SO

2

and O

3

; sense the intensity of

that fluorescence; generate various

electronic signals needed by the analyzer to

determine the SO

2

concentration of the sample gas (refer to Section 13.1.1), and sense

and control key environmental conditions such as the temperature of the sample gas and
the PMT.

Figure 13-15:

6200T Sensor Module

These components are divided into two significant subassemblies: the sample chamber
and the PMT assembly.

Figure 13-16 shows an exploded view of the sample chamber assembly
Figure 13-17shows an exploded view of the PMT Assembly

Advertising