6200T Total Sulfur Analyzer

Setup Menu

Teledyne Analytical Instruments

113

5.8. SETUP – VARS: Variables Setup and Definition

Through the SETUP>MORE>VARS menu there are several-user adjustable software
variables that define certain operational parameters. Usually, these variables are
automatically set by the instrument’s firmware, but can be manually re-defined using the
VARS menu. Table 5-2 lists all variables that are available within the 818 password
protected level.

Table 5-2: Variable Names (VARS) Revision G.3

NO.

VARIABLE

DESCRIPTION ALLOWED

VALUES

0

DAS_HOLD_OFF

Changes the internal data acquisition system (DAS) hold-off time,
which is the duration when data are not stored in the DAS because
the software considers the data to be questionable. That is the
case during warm-up or just after the instrument returns from one
of its calibration modes to SAMPLE mode. DAS_HOLD_OFF can
be disabled entirely in each DAS channel.

Can be between 0.5

and 20 minutes

Default=15 min.

1

TPC_ENABLE

Enables or disables the temperature and pressure compensation
(TPC) feature (refer to Section 13.7.3).

ON/OFF

2

RCELL_SET

Sets the sample chamber temperature. Increasing or decreasing
this temperature will increase or decrease the rate at which SO

2

*

decays into SO

2

(refer to Section 13.1.1).

Do not adjust this setting unless under the direction of Teledyne
Analytical Instruments customer service personnel.

30º C - 70º C

Default= 50º C

3

IZS_SET

Sets the IZS option temperature. Increasing or decreasing this
temperature will increase or decrease the permeation rate of the
IZS source (refer to Sections 3.3.2.4, 9.5, 9.6).

30º C - 70º C

Default= 50º C

4

DYN_ZERO

Dynamic zero automatically adjusts offset and slope of the SO

2

response when performing a zero point calibration during an
AutoCal (refer to Section 9).

ON/OFF

5

DYN_SPAN

Dynamic span automatically adjusts slope and slope of the SO

2

response when performing a zero point calibration during an
AutoCal (refer to Section 9).
Note that the DYN_ZERO and DYN_SPAN features are not
allowed for applications requiring EPA equivalency.

ON/OFF

6

CONC_PRECISION

Allows the user to set the number of significant digits to the right of
the decimal point display of concentration and stability values.

AUTO, 1, 2, 3, 4

Default=AUTO

7

CLOCK_ADJ

Adjusts the speed of the analyzer’s clock. Choose the + sign if the
clock is too slow, choose the - sign if the clock is too fast.

-60 to +60 s/day

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

There are more VARS available when using the password, 929, for
configuration. Use caution when pressing any buttons while in this setup.
Any changes made may alter the performance of the instrument or cause
the instrument to not function properly. Note that if there is an accidental
change to a setup parameter, press EXIT to discard the changes.