Aicom remote control program, Holdoff feature, 6200t total sulfur analyzer das and aicom – Teledyne 6200T - Sulfides Analyzer User Manual
Page 173
6200T Total Sulfur Analyzer
DAS and AICOM
Teledyne Analytical Instruments
173
7.2.11. HOLDOFF Feature
The DAS HOLDOFF feature allows to prevent data collection during calibrations and
during the DAS_HOLDOFF period enabled and specified in the VARS (refer to Section
6.8).
To enable or disable the HOLDOFF, follow the instruction shown in Figure 7-3, then
press:
Edit Data Channel Menu
SETUP X.X
0) CONC: ATIMER, 2, 4032, R
PREV NEXT INS DEL
EDIT
PRNT
EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
EXIT
SETUP X.X
CAL HOLD OFF:ON
SET>
EDIT
EXIT
ENTR
accepts the new
setting and returns to the
previous menu.
EXIT
ignores the new setting
and returns to the previous
menu.
SETUP X.X
CAL HOLD OFF:ON
ON
ENTR EXIT
Toggle to turn HOLDOFF
ON or OFF
Press SET> until…
Figure 7-12:
DAS – Holdoff Feature
The DAS can be configured and operated remotely via the AICOM program. Section
introduces AICOM and then describes how to
7.3. AICOM Remote Control Program
AICOM is an easy-to-use, yet powerful interface program that allows a user to access
and control any of Teledyne Analytical Instruments’s main line of ambient and stack-gas
instruments from a remote connection through direct cable, modem or Ethernet.
Running AICOM, a user can:
Establish a link from a remote location to the 6200T through direct cable
connection via RS-232 modem or Ethernet.
View the instrument’s front panel and remotely access all functions that
could be accessed when standing in front of the instrument.
Remotely edit system parameters and set points.
Download, view, graph and save data for predictive diagnostics or data
analysis.
Retrieve, view, edit, save and upload DAS configurations (Section 7.4).