Maintenance schedule, Maintenance model 6200t total sulfur analyzer, Teledyne analytical instruments 226 – Teledyne 6200T - Sulfides Analyzer User Manual

Page 226

Advertising
background image

Maintenance

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

226

11.1. Maintenance Schedule

Table 11-1 is the recommended maintenance schedule for the 6200T. Please note that in
certain environments with high levels of dust, humidity or pollutant levels some
maintenance procedures may need to be performed more often than shown.

Table 11-1:

6200T Preventive Maintenance Schedule

ITEM ACTION

FREQUENCY

CAL

CHECK

MANUAL

SECTION

DATE PERFORMED

1

Particulate filter

Change particle

filter

Weekly

No 11.3.1

Verify test functions

Review and

evaluate

Weekly No

11.2;

Appendix C

Zero/span check

Evaluate offset

and slope

Weekly

--

9.3, 9.6, 9.9

1

Zero/span calibration

Zero and span

calibration

Every 3 months

--

9.2, 9.4, 9.5,

9.7, 9.8

1

TS and zero air

scrubbers

Exchange

chemical

Every 3 months

No 11.3.3

1

Perform flow check

Check Flow

Every 6 Months

No

11.3.7

Internal IZS

Permeation Tube

Replace Annually YES

11.3.2

Perform pneumatic

leak check

Verify Leak

Tight

Annually or after repairs

involving pneumatics

Yes 11.3.6

2

Pump

diaphragm

Replace Annually Yes

Refer to

diaphragm kit

instructions

Calibrate UV Lamp

Output

Perform LAMP

CAL

Prior to zero/span

calibration or PMT

hardware calibration

--

5.9.6 &

12.7.2.5

3

PMT sensor

hardware calibration

Low-level
hardware

calibration

On PMT/ preamp

changes if

0.7 < SLOPE or

SLOPE >

1.3

Yes 12.7.2.8

1

Sample chamber

optics

Clean

chamber,

windows and

filters

As necessary

Yes

12.7.2.2 &

12.7.2.3

1

Critical flow orifice &

sintered filters

Replace As

necessary

Yes 11.3.4

1

These Items are required to maintain full warranty; all other items are strongly recommended.

2

A pump rebuild kit is available from Teledyne Analytical Instruments’s Customer Service including all instructions and required parts (refer to Appendix B for

3

Replace desiccant bags each time the inspection plate for the sensor assembly is removed.