Troubleshooting model 6200t total sulfur analyzer – Teledyne 6200T - Sulfides Analyzer User Manual

Page 240

Advertising
background image

Troubleshooting

Model 6200T Total Sulfur Analyzer

240

Teledyne Analytical Instruments

To view or clear a warning message press:

SAMPLE RANGE = 500.0 PPB SO2 =XXX.X

TEST

CAL

MSG

CLR

SETUP

If warning messages reappear,

the cause needs to be found. Do

not repeatedly clear warnings

without corrective action.

Press CLR to clear the current

warning message.

If more than one warning is

active, the next message will

take its place.

Once the last warning has been

cleared, the analyzer returns to

SAMPLE

Mode.

SAMPLE

RANGE = 500.0 PPB

SO2 =XXX.X


< TST TST > CAL

MSG

CLR

SETUP

SAMPLE

SYSTEM RESET

SO2= X.XXX


< TST TST > CAL

MSG

CLR

SETUP

In WARNING mode, <TST TST>

buttons replaced with TEST

button. Pressing TEST switches to
SAMPLE mode and hides warning

messages until new warning(s)

are activated.

MSG

indicates that one or more

warning message are active but

hidden. Pressing MSG cycles

through warnings

In SAMPLE mode, all warning

messages are hidden, but MSG

button appears

Figure 12-1:

Viewing and Clearing Warning Messages

Table 12-1:

Warning Messages - Indicated Failures

Warning Message

Fault Condition

Possible Causes

ANALOG CAL

WARNING

The instruments A/D

circuitry or one of its analog

outputs is not calibrated

A parameter for one of the analog outputs, even one not currently being used,

has been changed and the analog output calibration routine was not re-run

A/D circuitry failure on motherboard
Other motherboard electronic failure

BOX TEMP WARNING

Box Temp is < 5°C or >

48°C.

NOTE: Box temperature typically runs ~7

o

c warmer than ambient temperature.

Poor/blocked ventilation to the analyzer.
Stopped exhaust-fan
Ambient temperature outside of specified range

CANNOT DYN SPAN

Dynamic Span operation

failed

Measured concentration value is too high or low.
Concentration slope value too high or too low

CANNOT DYN ZERO

Dynamic Zero operation

failed

Measured concentration value is too high.
Concentration offset value too high.

CONFIG INITIALIZED

Configuration and

Calibration data reset to

original Factory state.

Failed disk on module
User erased data

DARK CAL WARNING

The Dark Cal signal is

higher than 200 mV.

Light leak in reaction cell
Shutter solenoid is not functioning
Failed relay board
I

2

C bus failure

Loose connector/wiring
PMT preamp board bad or out of cal

DATA INITIALIZED

Data Storage in DAS was

erased

Failed disk on module
User cleared data

HVPS WARNING

High voltage power supply

output is <400 V or >900 V

High voltage power supply is bad
High voltage power supply is out of cal
A/D converter circuitry is bad

IZS TEMP WARNING

On units with IZS options

installed: The permeation

tube temperature is Sample

chamber temperature is

< 45°C or > 55°C

Bad IZS heater
Bad IZS temperature sensor
Bad relay controlling the IZS heater
Entire relay board is malfunctioning
I

2

C bus malfunction

Failure of thermistor interface circuitry on motherboard

PMT DET WARNING

PMT detector output is >

4995 mV

Failed PMT
Malfunctioning PMR preamp board
A/D converter circuitry failure