12.6.10. Shutter System

To check the functionality of the UV light Shutter by manually activating it:

EXIT

4x’s to return

to the

SAMPLE

display

DIAG I / O

36) UVLAMP_SIGNAL= 3.4 MV

PREV NEXT

JUMP

PRNT

EXIT

SAMPLE RANGE = 500.000 PPB SO2 =XXX.X

< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

DIAG

SIGNAL I / O

PREV NEXT JUMP

ENTR

EXIT

DIAG I / O

0)

EXT_ZERO_CAL=OFF

PREV NEXT

JUMP

PRNT EXIT

Press these buttons

until 32 is displayed

DIAG I / O

JUMP TO: 01

0 1

ENTR

EXIT

Activate the

Dark Shutter

DIAG I / O

32) DARK_SHUTTER=OFF

PREV NEXT

JUMP

OFF

PRNT

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

EXIT

EXIT

returns

to the main

SAMPLE

display

DIAG I / O

32) DARK_SHUTTER=ON

PREV NEXT

JUMP

ON

PRNT

EXIT

DIAG I / O

JUMP TO: 32

3 2

ENTR

EXIT

Press these buttons

until 36 is displayed

UV LAMP_SIGNAL

should be

<20 mV

Figure 12-5:

Manual Activation of the UV Light Shutter

12.6.11. PMT Sensor

The photo multiplier tube detects the light emitted by the UV excited fluorescence of
SO

2

. It has a gain of about 500000 to 1000000. It is not possible to test the detector

outside of the instrument in the field. The best way to determine if the PMT is working
properly is by using the optical test (OTEST), which is described in Section 5.9.4. The
basic method to diagnose a PMT fault is to eliminate the other components using
ETEST, OTEST and specific tests for other sub-assemblies.

12.6.12. PMT Preamplifier Board

To check the correct operation of the preamplifier board, we suggest the technician carry
out the electrical and optical tests described in 5.9.4 and 5.9.5.
If the ETEST fails, the preamplifier board may be faulty.

12.6.13. PMT Temperature Control PCA

The TEC control printed circuit assembly is located on the sensor housing assembly,
under the slanted shroud, next to the cooling fins and directly above the cooling fan.

 If the red LED located on the top edge of this assembly is not glowing the

control circuit is not receiving power.