6200t total sulfur analyzer getting started, Teledyne analytical instruments 85, Figure 3-28: zero/span calibration procedure – Teledyne 6200T - Sulfides Analyzer User Manual
Page 85
6200T Total Sulfur Analyzer
Getting Started
Teledyne Analytical Instruments
85
Zero/Span Calibration
To perform the zero/span calibration procedure, press:
Press ENTR to changes
the OFFSET & SLOPE
values for the SO
2
measurements.
Press EXIT to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE
STABIL= XXXX PPB
SO2=XXX.X
< TST TST > CAL SETUP
M-P CAL
STABIL= XXXX PPB
SO2=XXX.X
<TST TST>
ZERO
CONC
EXIT
EXIT
at this point
returns to the
SAMPLE menu.
Press ENTR to changes
the OFFSET & SLOPE
values for the SO
2
measurements.
Press EXIT to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE RANGE=500.0 PPB SO2= XXXX
< TST
TST >
CAL SETUP
Set the Display to show
the STABIL test function.
This function calculates
the stability of the SO
2
measurement.
Toggle TST> button until ...
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until STABIL
falls below 0.5 ppb.
This may take several
minutes.
SAMPLE
STABIL= XXXX PPB
SO2=XXX.X
< TST TST >
CAL
SETUP
M-P CAL
STABIL= XXXX PPB
SO2=XXX.X
<TST TST>
ENTR
CONC
EXIT
M-P CAL
STABIL= XXXX PPB
SO2=XXX.X
<TST TST> ZERO
SPAN
CONC
EXIT
Allow span gas to enter the sample port
at the rear of the analyzer.
Wait until STABIL
falls below 0.5 ppb.
This may take several
minutes.
SAMPLE
STABIL= XXXX PPB
SO2=XXX.X
< TST TST >
CAL
SETUP
M-P CAL
STABIL= XXXX PPB
SO2=XXX.X
<TST TST>
ENTR
CONC
EXIT
M-P CAL
STABIL= XXXX PPB
SO2=XXX.X
<TST TST> ENTR
CONC
EXIT
The SPAN button now
appears during the transition
from zero to span.
You may see both buttons.
If either the ZERO or SPAN
buttons fail to appear see the
Troubleshooting section for
tips.
Figure 3-28:
Zero/Span Calibration Procedure