Zero/Span Calibration

To perform the zero/span calibration procedure, press:

Press ENTR to changes

the OFFSET & SLOPE

values for the SO

2

measurements.

Press EXIT to leave the

calibration unchanged and

return to the previous

menu.

SAMPLE

STABIL= XXXX PPB

SO2=XXX.X

< TST TST > CAL SETUP

M-P CAL

STABIL= XXXX PPB

SO2=XXX.X

<TST TST>

ZERO

CONC

EXIT

EXIT

at this point

returns to the

SAMPLE menu.

Press ENTR to changes

the OFFSET & SLOPE

values for the SO

2

measurements.

Press EXIT to leave the

calibration unchanged and

return to the previous

menu.

SAMPLE RANGE=500.0 PPB SO2= XXXX

< TST

TST >

CAL SETUP

Set the Display to show

the STABIL test function.

This function calculates

the stability of the SO

2

measurement.

Toggle TST> button until ...

Allow zero gas to enter the sample port

at the rear of the analyzer.

Wait until STABIL

falls below 0.5 ppb.

This may take several

minutes.

SAMPLE

STABIL= XXXX PPB

SO2=XXX.X

< TST TST >

CAL

SETUP

M-P CAL

STABIL= XXXX PPB

SO2=XXX.X

<TST TST>

ENTR

CONC

EXIT

M-P CAL

STABIL= XXXX PPB

SO2=XXX.X

<TST TST> ZERO

SPAN

CONC

EXIT

Allow span gas to enter the sample port

at the rear of the analyzer.

Wait until STABIL

falls below 0.5 ppb.

This may take several

minutes.

SAMPLE

STABIL= XXXX PPB

SO2=XXX.X

< TST TST >

CAL

SETUP

M-P CAL

STABIL= XXXX PPB

SO2=XXX.X

<TST TST>

ENTR

CONC

EXIT

M-P CAL

STABIL= XXXX PPB

SO2=XXX.X

<TST TST> ENTR

CONC

EXIT

The SPAN button now

appears during the transition

from zero to span.

You may see both buttons.

If either the ZERO or SPAN

buttons fail to appear see the

Troubleshooting section for

tips.

Figure 3-28:

Zero/Span Calibration Procedure