Configurations – Teledyne 6200T - Sulfides Analyzer User Manual

Page 27

Advertising
background image

6200T Total Sulfur Analyzer

Introduction

Teledyne Analytical Instruments

27

OPTION

OPTION

NUMBER

DESCRIPTION/NOTES REFERENCE

Hessen protocol networks.

Call Customer Service for activation.

N/A

Second Language Switch

activates an alternate set of display

messages in a language other than the instrument’s default language.

Call Customer Service for a specially programmed Disk on Module containing
the second language.

N/A

N/A

Dilution Ratio Option

allows the user to compensate for diluted

sample gas, such as in continuous emission monitoring (CEM) where
the quality of gas in a smoke stack is being tested and the sampling
method used to remove the gas from the stack dilutes the gas.

Call Customer Service for activation.

Section 3.4.4.1,
5.4.5 and 13.1.9.3

1.5. Configurations

There are three configurations available: the standard analyzer and two with options.
The standard unit consists of:

 A modified 6400T Fluorescent SO

2

Analyzer

 The M501TS High Temperature Thermal Converter
 External Span, Internal Zero with High-performance Charcoal Scrubber for

Zero.

See Figure 1-1 for the pneumatic diagram, and Section 1.6 for details on the M501TS.
Another configuration consists of:

 Standard unit (described above)
 Internal Zero/Span (IZS) Option with H

2

S permeation tube.

The IZS option uses sample gas (passed through a special, high-performance charcoal
scrubber) to dilute H

2

S from the perm tube for span calibration checks. See Figure 1-2

for the pneumatic diagram.
The third configuration consists of:

 Standard unit (described above)
 External M702 calibrator

The M702 calibrator option blends tanks of H

2

S span gas with the processed CO

2

.

See Figure 1-3 for the pneumatic diagram.

Advertising